Loading [MathJax]/extensions/tex2jax.js

Surface Roughness-Dependent Corrosion: Implications for Cochlear Implant Reliability

Research output: Contribution to journalArticleResearchpeer review

Details

Original languageEnglish
Pages (from-to)91-94
Number of pages4
JournalCurrent Directions in Biomedical Engineering
Volume10
Issue number4
Publication statusPublished - 19 Dec 2024

Abstract

The corrosion behavior of platinum electrodes in cochlear implant (CI) systems is a critical factor affecting their long-term functionality. This study investigated the influence of the surface roughness of the platinum electrodes on their corrosion behavior. The results of in vitro experiments indicate that higher surface roughness tends to accelerate corrosion by providing numerous initiation sites. Although the electrochemical measurements of platinum electrodes with different surface roughnesses showed only slight differences in terms of shifts in potentials and higher corrosion current densities, data from the literature suggests an enhanced charge capacity with rougher surfaces. These findings contribute to an understanding of the failure mechanisms of CIs and can ultimately help to improve the design and durability of CI systems.

Keywords

    cochlear implant, corrosion, cyclic voltammetry, platinum, polarization measurement, roughness

ASJC Scopus subject areas

Cite this

Surface Roughness-Dependent Corrosion: Implications for Cochlear Implant Reliability. / Blank, Tatiana; Behrens, Sabine; Klose, Christian et al.
In: Current Directions in Biomedical Engineering, Vol. 10, No. 4, 19.12.2024, p. 91-94.

Research output: Contribution to journalArticleResearchpeer review

Blank, T, Behrens, S, Klose, C & Maier, HJ 2024, 'Surface Roughness-Dependent Corrosion: Implications for Cochlear Implant Reliability', Current Directions in Biomedical Engineering, vol. 10, no. 4, pp. 91-94. https://doi.org/10.1515/cdbme-2024-2022
Blank T, Behrens S, Klose C, Maier HJ. Surface Roughness-Dependent Corrosion: Implications for Cochlear Implant Reliability. Current Directions in Biomedical Engineering. 2024 Dec 19;10(4):91-94. doi: 10.1515/cdbme-2024-2022
Blank, Tatiana ; Behrens, Sabine ; Klose, Christian et al. / Surface Roughness-Dependent Corrosion : Implications for Cochlear Implant Reliability. In: Current Directions in Biomedical Engineering. 2024 ; Vol. 10, No. 4. pp. 91-94.
Download
@article{48d631e022144d4db17fc4c080108d73,
title = "Surface Roughness-Dependent Corrosion: Implications for Cochlear Implant Reliability",
abstract = "The corrosion behavior of platinum electrodes in cochlear implant (CI) systems is a critical factor affecting their long-term functionality. This study investigated the influence of the surface roughness of the platinum electrodes on their corrosion behavior. The results of in vitro experiments indicate that higher surface roughness tends to accelerate corrosion by providing numerous initiation sites. Although the electrochemical measurements of platinum electrodes with different surface roughnesses showed only slight differences in terms of shifts in potentials and higher corrosion current densities, data from the literature suggests an enhanced charge capacity with rougher surfaces. These findings contribute to an understanding of the failure mechanisms of CIs and can ultimately help to improve the design and durability of CI systems.",
keywords = "cochlear implant, corrosion, cyclic voltammetry, platinum, polarization measurement, roughness",
author = "Tatiana Blank and Sabine Behrens and Christian Klose and Maier, {Hans J{\"u}rgen}",
note = "Publisher Copyright: {\textcopyright} 2024 by Walter de Gruyter Berlin/Boston.",
year = "2024",
month = dec,
day = "19",
doi = "10.1515/cdbme-2024-2022",
language = "English",
volume = "10",
pages = "91--94",
number = "4",

}

Download

TY - JOUR

T1 - Surface Roughness-Dependent Corrosion

T2 - Implications for Cochlear Implant Reliability

AU - Blank, Tatiana

AU - Behrens, Sabine

AU - Klose, Christian

AU - Maier, Hans Jürgen

N1 - Publisher Copyright: © 2024 by Walter de Gruyter Berlin/Boston.

PY - 2024/12/19

Y1 - 2024/12/19

N2 - The corrosion behavior of platinum electrodes in cochlear implant (CI) systems is a critical factor affecting their long-term functionality. This study investigated the influence of the surface roughness of the platinum electrodes on their corrosion behavior. The results of in vitro experiments indicate that higher surface roughness tends to accelerate corrosion by providing numerous initiation sites. Although the electrochemical measurements of platinum electrodes with different surface roughnesses showed only slight differences in terms of shifts in potentials and higher corrosion current densities, data from the literature suggests an enhanced charge capacity with rougher surfaces. These findings contribute to an understanding of the failure mechanisms of CIs and can ultimately help to improve the design and durability of CI systems.

AB - The corrosion behavior of platinum electrodes in cochlear implant (CI) systems is a critical factor affecting their long-term functionality. This study investigated the influence of the surface roughness of the platinum electrodes on their corrosion behavior. The results of in vitro experiments indicate that higher surface roughness tends to accelerate corrosion by providing numerous initiation sites. Although the electrochemical measurements of platinum electrodes with different surface roughnesses showed only slight differences in terms of shifts in potentials and higher corrosion current densities, data from the literature suggests an enhanced charge capacity with rougher surfaces. These findings contribute to an understanding of the failure mechanisms of CIs and can ultimately help to improve the design and durability of CI systems.

KW - cochlear implant

KW - corrosion

KW - cyclic voltammetry

KW - platinum

KW - polarization measurement

KW - roughness

UR - http://www.scopus.com/inward/record.url?scp=85218504784&partnerID=8YFLogxK

U2 - 10.1515/cdbme-2024-2022

DO - 10.1515/cdbme-2024-2022

M3 - Article

VL - 10

SP - 91

EP - 94

JO - Current Directions in Biomedical Engineering

JF - Current Directions in Biomedical Engineering

IS - 4

ER -

By the same author(s)