Repetition rate dependency of low-density plasma effects during femtosecond-laser-based surgery of biological tissue

Research output: Contribution to journalArticleResearchpeer review

Authors

  • Kai Kuetemeyer
  • Judith Baumgart
  • Holger Lubatschowski
  • Alexander Heisterkamp

External Research Organisations

  • Laser Zentrum Hannover e.V. (LZH)
View graph of relations

Details

Original languageEnglish
Pages (from-to)695-699
Number of pages5
JournalApplied Physics B: Lasers and Optics
Volume97
Issue number3
Publication statusPublished - 3 Oct 2009
Externally publishedYes

Abstract

Femtosecond laser based nanosurgery of biological tissue is usually done in two different regimes. Depending on the application, low kHz repetition rates above the optical breakdown threshold or high MHz repetition rates in the low-density plasma regime are used. In contrast to the well understood optical breakdown, mechanisms leading to dissection below this threshold are not well known due to the complexity of chemical effects with high numbers of interacting molecules. Furthermore, the laser repetition rate may influence their efficiency. In this paper, we present our study on low-density plasma effects in biological tissue depending on repetition rate by static exposure of porcine corneal stroma to femtosecond pulses. We observed a continuous increase of the laser-induced damage with decreasing repetition rate over two orders of magnitude at constant numbers of applied laser pulses or constant laser pulse energies. Therefore, low repetition rates in the kHz regime are advantageous to minimize the total delivered energy to biological tissue during femtosecond laser irradiation. However, due to frequent excessive damage in this regime directly above the threshold, MHz repetition rates are preferable to create nanometer-sized cuts in the low-density plasma regime.

ASJC Scopus subject areas

Cite this

Repetition rate dependency of low-density plasma effects during femtosecond-laser-based surgery of biological tissue. / Kuetemeyer, Kai; Baumgart, Judith; Lubatschowski, Holger et al.
In: Applied Physics B: Lasers and Optics, Vol. 97, No. 3, 03.10.2009, p. 695-699.

Research output: Contribution to journalArticleResearchpeer review

Kuetemeyer K, Baumgart J, Lubatschowski H, Heisterkamp A. Repetition rate dependency of low-density plasma effects during femtosecond-laser-based surgery of biological tissue. Applied Physics B: Lasers and Optics. 2009 Oct 3;97(3):695-699. doi: 10.1007/s00340-009-3746-y
Kuetemeyer, Kai ; Baumgart, Judith ; Lubatschowski, Holger et al. / Repetition rate dependency of low-density plasma effects during femtosecond-laser-based surgery of biological tissue. In: Applied Physics B: Lasers and Optics. 2009 ; Vol. 97, No. 3. pp. 695-699.
Download
@article{33e59f3f4a9443029153e3cca54207d2,
title = "Repetition rate dependency of low-density plasma effects during femtosecond-laser-based surgery of biological tissue",
abstract = "Femtosecond laser based nanosurgery of biological tissue is usually done in two different regimes. Depending on the application, low kHz repetition rates above the optical breakdown threshold or high MHz repetition rates in the low-density plasma regime are used. In contrast to the well understood optical breakdown, mechanisms leading to dissection below this threshold are not well known due to the complexity of chemical effects with high numbers of interacting molecules. Furthermore, the laser repetition rate may influence their efficiency. In this paper, we present our study on low-density plasma effects in biological tissue depending on repetition rate by static exposure of porcine corneal stroma to femtosecond pulses. We observed a continuous increase of the laser-induced damage with decreasing repetition rate over two orders of magnitude at constant numbers of applied laser pulses or constant laser pulse energies. Therefore, low repetition rates in the kHz regime are advantageous to minimize the total delivered energy to biological tissue during femtosecond laser irradiation. However, due to frequent excessive damage in this regime directly above the threshold, MHz repetition rates are preferable to create nanometer-sized cuts in the low-density plasma regime.",
author = "Kai Kuetemeyer and Judith Baumgart and Holger Lubatschowski and Alexander Heisterkamp",
year = "2009",
month = oct,
day = "3",
doi = "10.1007/s00340-009-3746-y",
language = "English",
volume = "97",
pages = "695--699",
journal = "Applied Physics B: Lasers and Optics",
issn = "0946-2171",
publisher = "Springer Verlag",
number = "3",

}

Download

TY - JOUR

T1 - Repetition rate dependency of low-density plasma effects during femtosecond-laser-based surgery of biological tissue

AU - Kuetemeyer, Kai

AU - Baumgart, Judith

AU - Lubatschowski, Holger

AU - Heisterkamp, Alexander

PY - 2009/10/3

Y1 - 2009/10/3

N2 - Femtosecond laser based nanosurgery of biological tissue is usually done in two different regimes. Depending on the application, low kHz repetition rates above the optical breakdown threshold or high MHz repetition rates in the low-density plasma regime are used. In contrast to the well understood optical breakdown, mechanisms leading to dissection below this threshold are not well known due to the complexity of chemical effects with high numbers of interacting molecules. Furthermore, the laser repetition rate may influence their efficiency. In this paper, we present our study on low-density plasma effects in biological tissue depending on repetition rate by static exposure of porcine corneal stroma to femtosecond pulses. We observed a continuous increase of the laser-induced damage with decreasing repetition rate over two orders of magnitude at constant numbers of applied laser pulses or constant laser pulse energies. Therefore, low repetition rates in the kHz regime are advantageous to minimize the total delivered energy to biological tissue during femtosecond laser irradiation. However, due to frequent excessive damage in this regime directly above the threshold, MHz repetition rates are preferable to create nanometer-sized cuts in the low-density plasma regime.

AB - Femtosecond laser based nanosurgery of biological tissue is usually done in two different regimes. Depending on the application, low kHz repetition rates above the optical breakdown threshold or high MHz repetition rates in the low-density plasma regime are used. In contrast to the well understood optical breakdown, mechanisms leading to dissection below this threshold are not well known due to the complexity of chemical effects with high numbers of interacting molecules. Furthermore, the laser repetition rate may influence their efficiency. In this paper, we present our study on low-density plasma effects in biological tissue depending on repetition rate by static exposure of porcine corneal stroma to femtosecond pulses. We observed a continuous increase of the laser-induced damage with decreasing repetition rate over two orders of magnitude at constant numbers of applied laser pulses or constant laser pulse energies. Therefore, low repetition rates in the kHz regime are advantageous to minimize the total delivered energy to biological tissue during femtosecond laser irradiation. However, due to frequent excessive damage in this regime directly above the threshold, MHz repetition rates are preferable to create nanometer-sized cuts in the low-density plasma regime.

UR - http://www.scopus.com/inward/record.url?scp=70350567217&partnerID=8YFLogxK

U2 - 10.1007/s00340-009-3746-y

DO - 10.1007/s00340-009-3746-y

M3 - Article

AN - SCOPUS:70350567217

VL - 97

SP - 695

EP - 699

JO - Applied Physics B: Lasers and Optics

JF - Applied Physics B: Lasers and Optics

SN - 0946-2171

IS - 3