Abstract
In this paper we present a miniaturized flame ionization detector and flame spectrometer fabricated using conventional micromachining technologies. The main component of both devices is a micro burner unit, which uses minimal oxyhydrogen to produce a stable miniature flame. The oxyhydrogen is generated at low energy consumption by a miniaturized electrolysis cell, which, can be operated by battery. Because of the low oxyhydrogen consumption and the minute scale of the burner unit and electrolyzer the oxyhydrogen is generated as-required, rather than stored as in conventional systems. Thus, there is no explosion hazard and the devices are not only made easily portable, but also safe. Furthermore, these systems possess sensitivity and selectivity that is comparable to conventional systems. Concentrations down to 1 ppm have been demonstrated with the micro flame ionization detector and a detection limit in the ppb range appears within reach. The micro flame spectrometer is undergoing initial development, but measurements based on atomic emission spectrometry demonstrate already a detection limit only 100-fold above levels observed in conventional systems.
ASJC Scopus subject areas
- Materials Science(all)
- Electronic, Optical and Magnetic Materials
- Physics and Astronomy(all)
- Instrumentation
- Physics and Astronomy(all)
- Condensed Matter Physics
- Materials Science(all)
- Surfaces, Coatings and Films
- Materials Science(all)
- Metals and Alloys
- Engineering(all)
- Electrical and Electronic Engineering
- Materials Science(all)
- Materials Chemistry
Sustainable Development Goals
Cite this
- Standard
- Harvard
- Apa
- Vancouver
- BibTeX
- RIS
In: Sensors and Actuators, B: Chemical, Vol. 63, No. 3, 05.05.2000, p. 159-166.
Research output: Contribution to journal › Article › Research › peer review
}
TY - JOUR
T1 - Micro flame ionization detector and micro flame spectrometer
AU - Zimmermann, S.
AU - Wischhusen, S.
AU - Müller, J.
N1 - Funding information: The research project “Flame Ionization detector and Flame Spectrometer in Microsystem Technology” is financially supported by the Deutsche Forschungsgesellschaft.
PY - 2000/5/5
Y1 - 2000/5/5
N2 - In this paper we present a miniaturized flame ionization detector and flame spectrometer fabricated using conventional micromachining technologies. The main component of both devices is a micro burner unit, which uses minimal oxyhydrogen to produce a stable miniature flame. The oxyhydrogen is generated at low energy consumption by a miniaturized electrolysis cell, which, can be operated by battery. Because of the low oxyhydrogen consumption and the minute scale of the burner unit and electrolyzer the oxyhydrogen is generated as-required, rather than stored as in conventional systems. Thus, there is no explosion hazard and the devices are not only made easily portable, but also safe. Furthermore, these systems possess sensitivity and selectivity that is comparable to conventional systems. Concentrations down to 1 ppm have been demonstrated with the micro flame ionization detector and a detection limit in the ppb range appears within reach. The micro flame spectrometer is undergoing initial development, but measurements based on atomic emission spectrometry demonstrate already a detection limit only 100-fold above levels observed in conventional systems.
AB - In this paper we present a miniaturized flame ionization detector and flame spectrometer fabricated using conventional micromachining technologies. The main component of both devices is a micro burner unit, which uses minimal oxyhydrogen to produce a stable miniature flame. The oxyhydrogen is generated at low energy consumption by a miniaturized electrolysis cell, which, can be operated by battery. Because of the low oxyhydrogen consumption and the minute scale of the burner unit and electrolyzer the oxyhydrogen is generated as-required, rather than stored as in conventional systems. Thus, there is no explosion hazard and the devices are not only made easily portable, but also safe. Furthermore, these systems possess sensitivity and selectivity that is comparable to conventional systems. Concentrations down to 1 ppm have been demonstrated with the micro flame ionization detector and a detection limit in the ppb range appears within reach. The micro flame spectrometer is undergoing initial development, but measurements based on atomic emission spectrometry demonstrate already a detection limit only 100-fold above levels observed in conventional systems.
UR - http://www.scopus.com/inward/record.url?scp=0033749877&partnerID=8YFLogxK
U2 - 10.1016/S0925-4005(00)00353-1
DO - 10.1016/S0925-4005(00)00353-1
M3 - Article
AN - SCOPUS:0033749877
VL - 63
SP - 159
EP - 166
JO - Sensors and Actuators, B: Chemical
JF - Sensors and Actuators, B: Chemical
SN - 0925-4005
IS - 3
ER -