High-Resolution High Kinetic Energy Ion Mobility Spectrometer Based on a Low-Discrimination Tristate Ion Shutter

Research output: Contribution to journalArticleResearchpeer review

Authors

View graph of relations

Details

Original languageEnglish
Pages (from-to)5603-5611
Number of pages9
JournalAnalytical chemistry
Volume90
Issue number9
Early online date6 Apr 2018
Publication statusPublished - 1 May 2018

Abstract

High kinetic energy ion mobility spectrometry (HiKE-IMS) allows for sensitive trace gas analysis within seconds, mitigating many disadvantages of standard ion mobility spectrometers through operation at reduced pressure and high electric field strengths. However, these advantages usually come at the cost of reduced resolving power, ranging from a maximum of 75 down to 50 at a reduced field strength of 120 Td for the original device. In this work, we present an extended theory for HiKE-IMS resolving power and a novel tristate ion shutter principle able to achieve initial ion packet widths of 1 μs without significant mobility discrimination. Such an ultrashort injection time allows for improving the resolving power of the HiKE-IMS to 140 for a wide range of reduced electric field strengths. With this resolving power, separating all ion species generated from a mixture of benzene, toluene, and xylene is possible. Furthermore, a resolving power of 140 is sufficient to partially separate isotopologues under high electric field strengths.

ASJC Scopus subject areas

Cite this

High-Resolution High Kinetic Energy Ion Mobility Spectrometer Based on a Low-Discrimination Tristate Ion Shutter. / Kirk, Ansgar T.; Grube, Denise; Kobelt, Tim et al.
In: Analytical chemistry, Vol. 90, No. 9, 01.05.2018, p. 5603-5611.

Research output: Contribution to journalArticleResearchpeer review

Kirk AT, Grube D, Kobelt T, Wendt C, Zimmermann S. High-Resolution High Kinetic Energy Ion Mobility Spectrometer Based on a Low-Discrimination Tristate Ion Shutter. Analytical chemistry. 2018 May 1;90(9):5603-5611. Epub 2018 Apr 6. doi: 10.1021/acs.analchem.7b04586
Kirk, Ansgar T. ; Grube, Denise ; Kobelt, Tim et al. / High-Resolution High Kinetic Energy Ion Mobility Spectrometer Based on a Low-Discrimination Tristate Ion Shutter. In: Analytical chemistry. 2018 ; Vol. 90, No. 9. pp. 5603-5611.
Download
@article{77906d708c7c4d168e0f8040b01bc18f,
title = "High-Resolution High Kinetic Energy Ion Mobility Spectrometer Based on a Low-Discrimination Tristate Ion Shutter",
abstract = "High kinetic energy ion mobility spectrometry (HiKE-IMS) allows for sensitive trace gas analysis within seconds, mitigating many disadvantages of standard ion mobility spectrometers through operation at reduced pressure and high electric field strengths. However, these advantages usually come at the cost of reduced resolving power, ranging from a maximum of 75 down to 50 at a reduced field strength of 120 Td for the original device. In this work, we present an extended theory for HiKE-IMS resolving power and a novel tristate ion shutter principle able to achieve initial ion packet widths of 1 μs without significant mobility discrimination. Such an ultrashort injection time allows for improving the resolving power of the HiKE-IMS to 140 for a wide range of reduced electric field strengths. With this resolving power, separating all ion species generated from a mixture of benzene, toluene, and xylene is possible. Furthermore, a resolving power of 140 is sufficient to partially separate isotopologues under high electric field strengths.",
author = "Kirk, {Ansgar T.} and Denise Grube and Tim Kobelt and Cornelius Wendt and Stefan Zimmermann",
note = "Copyright {\textcopyright} 2018 American Chemical Society",
year = "2018",
month = may,
day = "1",
doi = "10.1021/acs.analchem.7b04586",
language = "English",
volume = "90",
pages = "5603--5611",
journal = "Analytical chemistry",
issn = "0003-2700",
publisher = "American Chemical Society",
number = "9",

}

Download

TY - JOUR

T1 - High-Resolution High Kinetic Energy Ion Mobility Spectrometer Based on a Low-Discrimination Tristate Ion Shutter

AU - Kirk, Ansgar T.

AU - Grube, Denise

AU - Kobelt, Tim

AU - Wendt, Cornelius

AU - Zimmermann, Stefan

N1 - Copyright © 2018 American Chemical Society

PY - 2018/5/1

Y1 - 2018/5/1

N2 - High kinetic energy ion mobility spectrometry (HiKE-IMS) allows for sensitive trace gas analysis within seconds, mitigating many disadvantages of standard ion mobility spectrometers through operation at reduced pressure and high electric field strengths. However, these advantages usually come at the cost of reduced resolving power, ranging from a maximum of 75 down to 50 at a reduced field strength of 120 Td for the original device. In this work, we present an extended theory for HiKE-IMS resolving power and a novel tristate ion shutter principle able to achieve initial ion packet widths of 1 μs without significant mobility discrimination. Such an ultrashort injection time allows for improving the resolving power of the HiKE-IMS to 140 for a wide range of reduced electric field strengths. With this resolving power, separating all ion species generated from a mixture of benzene, toluene, and xylene is possible. Furthermore, a resolving power of 140 is sufficient to partially separate isotopologues under high electric field strengths.

AB - High kinetic energy ion mobility spectrometry (HiKE-IMS) allows for sensitive trace gas analysis within seconds, mitigating many disadvantages of standard ion mobility spectrometers through operation at reduced pressure and high electric field strengths. However, these advantages usually come at the cost of reduced resolving power, ranging from a maximum of 75 down to 50 at a reduced field strength of 120 Td for the original device. In this work, we present an extended theory for HiKE-IMS resolving power and a novel tristate ion shutter principle able to achieve initial ion packet widths of 1 μs without significant mobility discrimination. Such an ultrashort injection time allows for improving the resolving power of the HiKE-IMS to 140 for a wide range of reduced electric field strengths. With this resolving power, separating all ion species generated from a mixture of benzene, toluene, and xylene is possible. Furthermore, a resolving power of 140 is sufficient to partially separate isotopologues under high electric field strengths.

UR - http://www.scopus.com/inward/record.url?scp=85046464850&partnerID=8YFLogxK

U2 - 10.1021/acs.analchem.7b04586

DO - 10.1021/acs.analchem.7b04586

M3 - Article

C2 - 29624371

AN - SCOPUS:85046464850

VL - 90

SP - 5603

EP - 5611

JO - Analytical chemistry

JF - Analytical chemistry

SN - 0003-2700

IS - 9

ER -

By the same author(s)