Abstract

Process analytical technology (PAT) is a guide to improve process development in biotech industry. Optical sensors such as near and mid infrared spectrometers fulfill an essential part for PAT. NIRS and MIRS were investigated as non-invasive on line monitoring tools for animal cell cultivations in order to predict critical process parameters, like cell parameters as well as substrate and metabolite concentrations. Eight cultivations were performed with frequent sampling. Variances between cultivations were induced by spiking experiments with intent to break correlations between analytes; to keep causality of the models; and to increase model robustness. Calibration models were built for each analyte using partial least-squares regression method. Cultivations chosen for validation were not part of the calibration set. Glucose concentration, cell density and viability were predicted by NIRS with a root mean square error of prediction (RMSEP) of 0.36g/L, 3.9 106cells/mL and 3.62% respectively. Based on MIR spectra glucose and lactate concentrations were predicted with a RMSEP of 0.16 and 0.14g/L respectively. Results show that MIRS has higher accuracy regarding the prediction of single analytes. For prediction of a main course of a cultivation, NIRS is much better suited than MIRS.

Keywords

    CHO cell cultivation, Infrared spectroscopy, Multivariate data analysis, Online monitoring

ASJC Scopus subject areas

Cite this

Comparative study of non-invasive monitoring via infrared spectroscopy for mammalian cell cultivations. / Sandor, M.; Rüdinger, F.; Bienert, R. et al.
In: Journal of biotechnology, Vol. 168, No. 4, 12.08.2013, p. 636-645.

Research output: Contribution to journalArticleResearchpeer review

Sandor M, Rüdinger F, Bienert R, Grimm C, Solle D, Scheper T. Comparative study of non-invasive monitoring via infrared spectroscopy for mammalian cell cultivations. Journal of biotechnology. 2013 Aug 12;168(4):636-645. doi: 10.1016/j.jbiotec.2013.08.002
Sandor, M. ; Rüdinger, F. ; Bienert, R. et al. / Comparative study of non-invasive monitoring via infrared spectroscopy for mammalian cell cultivations. In: Journal of biotechnology. 2013 ; Vol. 168, No. 4. pp. 636-645.
Download
@article{a2b316e410aa49b1b09e2e433469e8ff,
title = "Comparative study of non-invasive monitoring via infrared spectroscopy for mammalian cell cultivations",
abstract = "Process analytical technology (PAT) is a guide to improve process development in biotech industry. Optical sensors such as near and mid infrared spectrometers fulfill an essential part for PAT. NIRS and MIRS were investigated as non-invasive on line monitoring tools for animal cell cultivations in order to predict critical process parameters, like cell parameters as well as substrate and metabolite concentrations. Eight cultivations were performed with frequent sampling. Variances between cultivations were induced by spiking experiments with intent to break correlations between analytes; to keep causality of the models; and to increase model robustness. Calibration models were built for each analyte using partial least-squares regression method. Cultivations chosen for validation were not part of the calibration set. Glucose concentration, cell density and viability were predicted by NIRS with a root mean square error of prediction (RMSEP) of 0.36g/L, 3.9 106cells/mL and 3.62% respectively. Based on MIR spectra glucose and lactate concentrations were predicted with a RMSEP of 0.16 and 0.14g/L respectively. Results show that MIRS has higher accuracy regarding the prediction of single analytes. For prediction of a main course of a cultivation, NIRS is much better suited than MIRS.",
keywords = "CHO cell cultivation, Infrared spectroscopy, Multivariate data analysis, Online monitoring",
author = "M. Sandor and F. R{\"u}dinger and R. Bienert and C. Grimm and D. Solle and T. Scheper",
note = "Funding information: The authors would like to thank Sartorius Stedim Biotech GmbH for financial and technical support.",
year = "2013",
month = aug,
day = "12",
doi = "10.1016/j.jbiotec.2013.08.002",
language = "English",
volume = "168",
pages = "636--645",
journal = "Journal of biotechnology",
issn = "0168-1656",
publisher = "Elsevier",
number = "4",

}

Download

TY - JOUR

T1 - Comparative study of non-invasive monitoring via infrared spectroscopy for mammalian cell cultivations

AU - Sandor, M.

AU - Rüdinger, F.

AU - Bienert, R.

AU - Grimm, C.

AU - Solle, D.

AU - Scheper, T.

N1 - Funding information: The authors would like to thank Sartorius Stedim Biotech GmbH for financial and technical support.

PY - 2013/8/12

Y1 - 2013/8/12

N2 - Process analytical technology (PAT) is a guide to improve process development in biotech industry. Optical sensors such as near and mid infrared spectrometers fulfill an essential part for PAT. NIRS and MIRS were investigated as non-invasive on line monitoring tools for animal cell cultivations in order to predict critical process parameters, like cell parameters as well as substrate and metabolite concentrations. Eight cultivations were performed with frequent sampling. Variances between cultivations were induced by spiking experiments with intent to break correlations between analytes; to keep causality of the models; and to increase model robustness. Calibration models were built for each analyte using partial least-squares regression method. Cultivations chosen for validation were not part of the calibration set. Glucose concentration, cell density and viability were predicted by NIRS with a root mean square error of prediction (RMSEP) of 0.36g/L, 3.9 106cells/mL and 3.62% respectively. Based on MIR spectra glucose and lactate concentrations were predicted with a RMSEP of 0.16 and 0.14g/L respectively. Results show that MIRS has higher accuracy regarding the prediction of single analytes. For prediction of a main course of a cultivation, NIRS is much better suited than MIRS.

AB - Process analytical technology (PAT) is a guide to improve process development in biotech industry. Optical sensors such as near and mid infrared spectrometers fulfill an essential part for PAT. NIRS and MIRS were investigated as non-invasive on line monitoring tools for animal cell cultivations in order to predict critical process parameters, like cell parameters as well as substrate and metabolite concentrations. Eight cultivations were performed with frequent sampling. Variances between cultivations were induced by spiking experiments with intent to break correlations between analytes; to keep causality of the models; and to increase model robustness. Calibration models were built for each analyte using partial least-squares regression method. Cultivations chosen for validation were not part of the calibration set. Glucose concentration, cell density and viability were predicted by NIRS with a root mean square error of prediction (RMSEP) of 0.36g/L, 3.9 106cells/mL and 3.62% respectively. Based on MIR spectra glucose and lactate concentrations were predicted with a RMSEP of 0.16 and 0.14g/L respectively. Results show that MIRS has higher accuracy regarding the prediction of single analytes. For prediction of a main course of a cultivation, NIRS is much better suited than MIRS.

KW - CHO cell cultivation

KW - Infrared spectroscopy

KW - Multivariate data analysis

KW - Online monitoring

UR - http://www.scopus.com/inward/record.url?scp=84888824939&partnerID=8YFLogxK

U2 - 10.1016/j.jbiotec.2013.08.002

DO - 10.1016/j.jbiotec.2013.08.002

M3 - Article

C2 - 23948256

AN - SCOPUS:84888824939

VL - 168

SP - 636

EP - 645

JO - Journal of biotechnology

JF - Journal of biotechnology

SN - 0168-1656

IS - 4

ER -

By the same author(s)