A simple electrical circuit model for impedance spectroscopy with cochlear implant electrodes

Research output: Contribution to journalArticleResearchpeer review

Authors

External Research Organisations

  • Hannover Medical School (MHH)
View graph of relations

Details

Original languageEnglish
Article number109125
JournalHearing research
Volume453
Early online date4 Oct 2024
Publication statusPublished - Nov 2024

Abstract

Although cochlear implants are an established method of restoring hearing, they can have limitations such as increasing current spread and decreasing frequency resolution due to tissue growth around the electrode array. Impedance measurements in cochlear implants have become a versatile tool for intra- and post-operative diagnosis of cochlear implant state. However, most clinical devices use current pulse stimulation already available in the implants and analyze the voltage response in the time-domain and spread along the cochlea. To use the full potential of impedance spectroscopy in differentiating cell types, measurement over an extended frequency range is required. This study presents a simple electrical equivalent circuit for impedance spectroscopy with cochlear implants in a 2-pole configuration. The electrical equivalent circuit describes the electrical properties of the cochlear implant electrode and its electrochemical behavior at the electrode-electrolyte interface by comparing two non-linear bilayer models, Cole-Cole and Schwan-Faraday. The model is validated for four cochlear implant electrodes from four different manufacturers (MED-EL FlexSoft, AB HiFocus SlimJ, Oticon EVO, Cochlear Nucleus CI622) characterized by impedance spectroscopy between 5 Hz and 13 MHz. In the future, this electrical equivalent circuit may help to extract parameters for differentiating cell types around the cochlear implant electrode from an impedance spectroscopic measurement.

Keywords

    Cochlear implant, Electrical equivalent circuit, Electrode-electrolyte interface, Impedance spectroscopy

ASJC Scopus subject areas

Cite this

A simple electrical circuit model for impedance spectroscopy with cochlear implant electrodes. / Sehlmeyer, Merle; Bhavsar, Mit B.; Zimmermann, Stefan et al.
In: Hearing research, Vol. 453, 109125, 11.2024.

Research output: Contribution to journalArticleResearchpeer review

Sehlmeyer M, Bhavsar MB, Zimmermann S, Maier H. A simple electrical circuit model for impedance spectroscopy with cochlear implant electrodes. Hearing research. 2024 Nov;453:109125. Epub 2024 Oct 4. doi: 10.1016/j.heares.2024.109125
Download
@article{617e394236874df09bbf1afd2d8bd1b3,
title = "A simple electrical circuit model for impedance spectroscopy with cochlear implant electrodes",
abstract = "Although cochlear implants are an established method of restoring hearing, they can have limitations such as increasing current spread and decreasing frequency resolution due to tissue growth around the electrode array. Impedance measurements in cochlear implants have become a versatile tool for intra- and post-operative diagnosis of cochlear implant state. However, most clinical devices use current pulse stimulation already available in the implants and analyze the voltage response in the time-domain and spread along the cochlea. To use the full potential of impedance spectroscopy in differentiating cell types, measurement over an extended frequency range is required. This study presents a simple electrical equivalent circuit for impedance spectroscopy with cochlear implants in a 2-pole configuration. The electrical equivalent circuit describes the electrical properties of the cochlear implant electrode and its electrochemical behavior at the electrode-electrolyte interface by comparing two non-linear bilayer models, Cole-Cole and Schwan-Faraday. The model is validated for four cochlear implant electrodes from four different manufacturers (MED-EL FlexSoft, AB HiFocus SlimJ, Oticon EVO, Cochlear Nucleus CI622) characterized by impedance spectroscopy between 5 Hz and 13 MHz. In the future, this electrical equivalent circuit may help to extract parameters for differentiating cell types around the cochlear implant electrode from an impedance spectroscopic measurement.",
keywords = "Cochlear implant, Electrical equivalent circuit, Electrode-electrolyte interface, Impedance spectroscopy",
author = "Merle Sehlmeyer and Bhavsar, {Mit B.} and Stefan Zimmermann and Hannes Maier",
note = "Publisher Copyright: {\textcopyright} 2024 The Authors",
year = "2024",
month = nov,
doi = "10.1016/j.heares.2024.109125",
language = "English",
volume = "453",
journal = "Hearing research",
issn = "0378-5955",
publisher = "Elsevier",

}

Download

TY - JOUR

T1 - A simple electrical circuit model for impedance spectroscopy with cochlear implant electrodes

AU - Sehlmeyer, Merle

AU - Bhavsar, Mit B.

AU - Zimmermann, Stefan

AU - Maier, Hannes

N1 - Publisher Copyright: © 2024 The Authors

PY - 2024/11

Y1 - 2024/11

N2 - Although cochlear implants are an established method of restoring hearing, they can have limitations such as increasing current spread and decreasing frequency resolution due to tissue growth around the electrode array. Impedance measurements in cochlear implants have become a versatile tool for intra- and post-operative diagnosis of cochlear implant state. However, most clinical devices use current pulse stimulation already available in the implants and analyze the voltage response in the time-domain and spread along the cochlea. To use the full potential of impedance spectroscopy in differentiating cell types, measurement over an extended frequency range is required. This study presents a simple electrical equivalent circuit for impedance spectroscopy with cochlear implants in a 2-pole configuration. The electrical equivalent circuit describes the electrical properties of the cochlear implant electrode and its electrochemical behavior at the electrode-electrolyte interface by comparing two non-linear bilayer models, Cole-Cole and Schwan-Faraday. The model is validated for four cochlear implant electrodes from four different manufacturers (MED-EL FlexSoft, AB HiFocus SlimJ, Oticon EVO, Cochlear Nucleus CI622) characterized by impedance spectroscopy between 5 Hz and 13 MHz. In the future, this electrical equivalent circuit may help to extract parameters for differentiating cell types around the cochlear implant electrode from an impedance spectroscopic measurement.

AB - Although cochlear implants are an established method of restoring hearing, they can have limitations such as increasing current spread and decreasing frequency resolution due to tissue growth around the electrode array. Impedance measurements in cochlear implants have become a versatile tool for intra- and post-operative diagnosis of cochlear implant state. However, most clinical devices use current pulse stimulation already available in the implants and analyze the voltage response in the time-domain and spread along the cochlea. To use the full potential of impedance spectroscopy in differentiating cell types, measurement over an extended frequency range is required. This study presents a simple electrical equivalent circuit for impedance spectroscopy with cochlear implants in a 2-pole configuration. The electrical equivalent circuit describes the electrical properties of the cochlear implant electrode and its electrochemical behavior at the electrode-electrolyte interface by comparing two non-linear bilayer models, Cole-Cole and Schwan-Faraday. The model is validated for four cochlear implant electrodes from four different manufacturers (MED-EL FlexSoft, AB HiFocus SlimJ, Oticon EVO, Cochlear Nucleus CI622) characterized by impedance spectroscopy between 5 Hz and 13 MHz. In the future, this electrical equivalent circuit may help to extract parameters for differentiating cell types around the cochlear implant electrode from an impedance spectroscopic measurement.

KW - Cochlear implant

KW - Electrical equivalent circuit

KW - Electrode-electrolyte interface

KW - Impedance spectroscopy

UR - http://www.scopus.com/inward/record.url?scp=85206082236&partnerID=8YFLogxK

U2 - 10.1016/j.heares.2024.109125

DO - 10.1016/j.heares.2024.109125

M3 - Article

C2 - 39396445

AN - SCOPUS:85206082236

VL - 453

JO - Hearing research

JF - Hearing research

SN - 0378-5955

M1 - 109125

ER -

By the same author(s)