A compact high resolution ion mobility spectrometer for fast trace gas analysis

Research output: Contribution to journalArticleResearchpeer review

Authors

  • Ansgar T. Kirk
  • Maria Allers
  • Philipp Cochems
  • Jens Langejuergen
  • Stefan Zimmermann
Plum Print visual indicator of research metrics
  • Citations
    • Citation Indexes: 101
  • Captures
    • Readers: 66
  • Mentions
    • References: 2
see details

Details

Original languageEnglish
Pages (from-to)5200-5207
Number of pages8
JournalANALYST
Volume138
Issue number18
Publication statusPublished - 23 Apr 2013

Abstract

Drift tube ion mobility spectrometers (IMS) are widely used for fast trace gas detection in air, but portable compact systems are typically very limited in their resolving power. Decreasing the initial ion packet width improves the resolution, but is generally associated with a reduced signal-to-noise-ratio (SNR) due to the lower number of ions injected into the drift region. In this paper, we present a refined theory of IMS operation which employs a combined approach for the analysis of the ion drift and the subsequent amplification to predict both the resolution and the SNR of the measured ion current peak. This theoretical analysis shows that the SNR is not a function of the initial ion packet width, meaning that compact drift tube IMS with both very high resolution and extremely low limits of detection can be designed. Based on these implications, an optimized combination of a compact drift tube with a length of just 10 cm and a transimpedance amplifier has been constructed with a resolution of 183 measured for the positive reactant ion peak (RIP+), which is sufficient to e.g. separate the RIP+ from the protonated acetone monomer, even though their drift times only differ by a factor of 1.007. Furthermore, the limits of detection (LODs) for acetone are 180 pptv within 1 s of averaging time and 580 pptv within only 100 ms.

ASJC Scopus subject areas

Cite this

A compact high resolution ion mobility spectrometer for fast trace gas analysis. / Kirk, Ansgar T.; Allers, Maria; Cochems, Philipp et al.
In: ANALYST, Vol. 138, No. 18, 23.04.2013, p. 5200-5207.

Research output: Contribution to journalArticleResearchpeer review

Kirk AT, Allers M, Cochems P, Langejuergen J, Zimmermann S. A compact high resolution ion mobility spectrometer for fast trace gas analysis. ANALYST. 2013 Apr 23;138(18):5200-5207. doi: 10.1039/c3an00231d, 10.15488/63
Kirk, Ansgar T. ; Allers, Maria ; Cochems, Philipp et al. / A compact high resolution ion mobility spectrometer for fast trace gas analysis. In: ANALYST. 2013 ; Vol. 138, No. 18. pp. 5200-5207.
Download
@article{58fe61e35b344d7f8790bcb5eb0eacf5,
title = "A compact high resolution ion mobility spectrometer for fast trace gas analysis",
abstract = "Drift tube ion mobility spectrometers (IMS) are widely used for fast trace gas detection in air, but portable compact systems are typically very limited in their resolving power. Decreasing the initial ion packet width improves the resolution, but is generally associated with a reduced signal-to-noise-ratio (SNR) due to the lower number of ions injected into the drift region. In this paper, we present a refined theory of IMS operation which employs a combined approach for the analysis of the ion drift and the subsequent amplification to predict both the resolution and the SNR of the measured ion current peak. This theoretical analysis shows that the SNR is not a function of the initial ion packet width, meaning that compact drift tube IMS with both very high resolution and extremely low limits of detection can be designed. Based on these implications, an optimized combination of a compact drift tube with a length of just 10 cm and a transimpedance amplifier has been constructed with a resolution of 183 measured for the positive reactant ion peak (RIP+), which is sufficient to e.g. separate the RIP+ from the protonated acetone monomer, even though their drift times only differ by a factor of 1.007. Furthermore, the limits of detection (LODs) for acetone are 180 pptv within 1 s of averaging time and 580 pptv within only 100 ms.",
author = "Kirk, {Ansgar T.} and Maria Allers and Philipp Cochems and Jens Langejuergen and Stefan Zimmermann",
year = "2013",
month = apr,
day = "23",
doi = "10.1039/c3an00231d",
language = "English",
volume = "138",
pages = "5200--5207",
journal = "ANALYST",
issn = "0003-2654",
publisher = "Royal Society of Chemistry",
number = "18",

}

Download

TY - JOUR

T1 - A compact high resolution ion mobility spectrometer for fast trace gas analysis

AU - Kirk, Ansgar T.

AU - Allers, Maria

AU - Cochems, Philipp

AU - Langejuergen, Jens

AU - Zimmermann, Stefan

PY - 2013/4/23

Y1 - 2013/4/23

N2 - Drift tube ion mobility spectrometers (IMS) are widely used for fast trace gas detection in air, but portable compact systems are typically very limited in their resolving power. Decreasing the initial ion packet width improves the resolution, but is generally associated with a reduced signal-to-noise-ratio (SNR) due to the lower number of ions injected into the drift region. In this paper, we present a refined theory of IMS operation which employs a combined approach for the analysis of the ion drift and the subsequent amplification to predict both the resolution and the SNR of the measured ion current peak. This theoretical analysis shows that the SNR is not a function of the initial ion packet width, meaning that compact drift tube IMS with both very high resolution and extremely low limits of detection can be designed. Based on these implications, an optimized combination of a compact drift tube with a length of just 10 cm and a transimpedance amplifier has been constructed with a resolution of 183 measured for the positive reactant ion peak (RIP+), which is sufficient to e.g. separate the RIP+ from the protonated acetone monomer, even though their drift times only differ by a factor of 1.007. Furthermore, the limits of detection (LODs) for acetone are 180 pptv within 1 s of averaging time and 580 pptv within only 100 ms.

AB - Drift tube ion mobility spectrometers (IMS) are widely used for fast trace gas detection in air, but portable compact systems are typically very limited in their resolving power. Decreasing the initial ion packet width improves the resolution, but is generally associated with a reduced signal-to-noise-ratio (SNR) due to the lower number of ions injected into the drift region. In this paper, we present a refined theory of IMS operation which employs a combined approach for the analysis of the ion drift and the subsequent amplification to predict both the resolution and the SNR of the measured ion current peak. This theoretical analysis shows that the SNR is not a function of the initial ion packet width, meaning that compact drift tube IMS with both very high resolution and extremely low limits of detection can be designed. Based on these implications, an optimized combination of a compact drift tube with a length of just 10 cm and a transimpedance amplifier has been constructed with a resolution of 183 measured for the positive reactant ion peak (RIP+), which is sufficient to e.g. separate the RIP+ from the protonated acetone monomer, even though their drift times only differ by a factor of 1.007. Furthermore, the limits of detection (LODs) for acetone are 180 pptv within 1 s of averaging time and 580 pptv within only 100 ms.

UR - http://www.scopus.com/inward/record.url?scp=84881633547&partnerID=8YFLogxK

U2 - 10.1039/c3an00231d

DO - 10.1039/c3an00231d

M3 - Article

AN - SCOPUS:84881633547

VL - 138

SP - 5200

EP - 5207

JO - ANALYST

JF - ANALYST

SN - 0003-2654

IS - 18

ER -

By the same author(s)