Trace element diffusion in rhyolitic melts: Comparison between synchrotron radiation X-ray fluorescene microanalysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS)

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Autoren

  • Matthias Hahn
  • Harald Behrens
  • Astrid Tegge-Schüring
  • Jürgen Koepke
  • Ingo Horn
  • Karen Rickers
  • Gerald Falkenberg
  • Michael Wiedenbeck

Organisationseinheiten

Externe Organisationen

  • Deutsches Elektronen-Synchrotron (DESY)
  • Helmholtz-Zentrum Potsdam Deutsches GeoForschungsZentrum (GFZ)
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Seiten (von - bis)233-242
Seitenumfang10
FachzeitschriftEuropean journal of mineralogy
Jahrgang17
Ausgabenummer2
PublikationsstatusVeröffentlicht - 29 Apr. 2005

Abstract

Two microbeam techniques, synchrotron radiation X-ray fluorescence micro-analysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS) are compared for analyzing diffusion profiles of trace elements in two hydrous rhyolitic glasses (1.87 and 5.00wt% H2O). In order to verify the results, laser ablation coupled to inductively coupled plasma optical emission (LA-ICP-OES) has been used on one sample. Samples were produced by diffusion couple experiments performed in an internally heated gas pressure vessel at 1200°C and 500 MPa. One half of each couple was doped with 24 trace elements representing different geochemical groups: low field strength elements (Rb, Sr, Ba), transition metals (Cr, Co, Ni, Cu, Zn), rare earth elements (La, Ce, Nd, Sm, Eu, Gd, Er, Yb) + Y, high field strength elements (V, Zr, Nb, Hf, Ta) and main group elements (Ge, Sn). Several profiles were measured with both p-SRXRF and SIMS on both samples. In principle, concentrations of all elements can be extracted simultaneously from a single SRXRF spectrum. However, some trace elements could not be reliably quantified with our analytical system: Ta and Pb (used for detector collimator material), Ti, V (low energy of Kα, Co (Kα-peak overlapping with Fe K β-peak) and Cr, Ni, Cu, Zn (overlapping with 1-lines of REEs). In contrast, SIMS analyses measure each element sequentially. Hence, not all elements of the large total set of trace elements could be analyzed in a single run. Some elements requiring a high mass resolution (NaSi interfering with V, CaO interfering with Ni) or having low yields (Sn) were not profiled. Multi ple diffusivities derived from V-SRXRF and SIMS profiles are in very good agreement for most elements. In general, the trace element diffusivity decreases with increasing valence state, e.g. in sample D22 containing 1.87 wt% H 2O from log D = -10.80 for the monovalent Rb to log D=-13.34 for the tetravalent Zr (Din m2/s). By increasing the water content in sample D18 to 5.00 wt%, diffusion coefficients increase approximately by one order of magnitude for all elements studied.

ASJC Scopus Sachgebiete

Zitieren

Trace element diffusion in rhyolitic melts: Comparison between synchrotron radiation X-ray fluorescene microanalysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS). / Hahn, Matthias; Behrens, Harald; Tegge-Schüring, Astrid et al.
in: European journal of mineralogy, Jahrgang 17, Nr. 2, 29.04.2005, S. 233-242.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Hahn M, Behrens H, Tegge-Schüring A, Koepke J, Horn I, Rickers K et al. Trace element diffusion in rhyolitic melts: Comparison between synchrotron radiation X-ray fluorescene microanalysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS). European journal of mineralogy. 2005 Apr 29;17(2):233-242. doi: 10.1127/0935-1221/2005/0017-0233
Download
@article{b1106c22743d4158bffe76e6cdafc7da,
title = "Trace element diffusion in rhyolitic melts: Comparison between synchrotron radiation X-ray fluorescene microanalysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS)",
abstract = "Two microbeam techniques, synchrotron radiation X-ray fluorescence micro-analysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS) are compared for analyzing diffusion profiles of trace elements in two hydrous rhyolitic glasses (1.87 and 5.00wt% H2O). In order to verify the results, laser ablation coupled to inductively coupled plasma optical emission (LA-ICP-OES) has been used on one sample. Samples were produced by diffusion couple experiments performed in an internally heated gas pressure vessel at 1200°C and 500 MPa. One half of each couple was doped with 24 trace elements representing different geochemical groups: low field strength elements (Rb, Sr, Ba), transition metals (Cr, Co, Ni, Cu, Zn), rare earth elements (La, Ce, Nd, Sm, Eu, Gd, Er, Yb) + Y, high field strength elements (V, Zr, Nb, Hf, Ta) and main group elements (Ge, Sn). Several profiles were measured with both p-SRXRF and SIMS on both samples. In principle, concentrations of all elements can be extracted simultaneously from a single SRXRF spectrum. However, some trace elements could not be reliably quantified with our analytical system: Ta and Pb (used for detector collimator material), Ti, V (low energy of Kα, Co (Kα-peak overlapping with Fe K β-peak) and Cr, Ni, Cu, Zn (overlapping with 1-lines of REEs). In contrast, SIMS analyses measure each element sequentially. Hence, not all elements of the large total set of trace elements could be analyzed in a single run. Some elements requiring a high mass resolution (NaSi interfering with V, CaO interfering with Ni) or having low yields (Sn) were not profiled. Multi ple diffusivities derived from V-SRXRF and SIMS profiles are in very good agreement for most elements. In general, the trace element diffusivity decreases with increasing valence state, e.g. in sample D22 containing 1.87 wt% H 2O from log D = -10.80 for the monovalent Rb to log D=-13.34 for the tetravalent Zr (Din m2/s). By increasing the water content in sample D18 to 5.00 wt%, diffusion coefficients increase approximately by one order of magnitude for all elements studied.",
keywords = "Rhyolite, SIMS, Synchroton radiation X-ray fluorescence microanalysis, Trace element diffusion",
author = "Matthias Hahn and Harald Behrens and Astrid Tegge-Sch{\"u}ring and J{\"u}rgen Koepke and Ingo Horn and Karen Rickers and Gerald Falkenberg and Michael Wiedenbeck",
year = "2005",
month = apr,
day = "29",
doi = "10.1127/0935-1221/2005/0017-0233",
language = "English",
volume = "17",
pages = "233--242",
journal = "European journal of mineralogy",
issn = "0935-1221",
publisher = "E. Schweizerbart'sche Verlagsbuchhandlung",
number = "2",

}

Download

TY - JOUR

T1 - Trace element diffusion in rhyolitic melts

T2 - Comparison between synchrotron radiation X-ray fluorescene microanalysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS)

AU - Hahn, Matthias

AU - Behrens, Harald

AU - Tegge-Schüring, Astrid

AU - Koepke, Jürgen

AU - Horn, Ingo

AU - Rickers, Karen

AU - Falkenberg, Gerald

AU - Wiedenbeck, Michael

PY - 2005/4/29

Y1 - 2005/4/29

N2 - Two microbeam techniques, synchrotron radiation X-ray fluorescence micro-analysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS) are compared for analyzing diffusion profiles of trace elements in two hydrous rhyolitic glasses (1.87 and 5.00wt% H2O). In order to verify the results, laser ablation coupled to inductively coupled plasma optical emission (LA-ICP-OES) has been used on one sample. Samples were produced by diffusion couple experiments performed in an internally heated gas pressure vessel at 1200°C and 500 MPa. One half of each couple was doped with 24 trace elements representing different geochemical groups: low field strength elements (Rb, Sr, Ba), transition metals (Cr, Co, Ni, Cu, Zn), rare earth elements (La, Ce, Nd, Sm, Eu, Gd, Er, Yb) + Y, high field strength elements (V, Zr, Nb, Hf, Ta) and main group elements (Ge, Sn). Several profiles were measured with both p-SRXRF and SIMS on both samples. In principle, concentrations of all elements can be extracted simultaneously from a single SRXRF spectrum. However, some trace elements could not be reliably quantified with our analytical system: Ta and Pb (used for detector collimator material), Ti, V (low energy of Kα, Co (Kα-peak overlapping with Fe K β-peak) and Cr, Ni, Cu, Zn (overlapping with 1-lines of REEs). In contrast, SIMS analyses measure each element sequentially. Hence, not all elements of the large total set of trace elements could be analyzed in a single run. Some elements requiring a high mass resolution (NaSi interfering with V, CaO interfering with Ni) or having low yields (Sn) were not profiled. Multi ple diffusivities derived from V-SRXRF and SIMS profiles are in very good agreement for most elements. In general, the trace element diffusivity decreases with increasing valence state, e.g. in sample D22 containing 1.87 wt% H 2O from log D = -10.80 for the monovalent Rb to log D=-13.34 for the tetravalent Zr (Din m2/s). By increasing the water content in sample D18 to 5.00 wt%, diffusion coefficients increase approximately by one order of magnitude for all elements studied.

AB - Two microbeam techniques, synchrotron radiation X-ray fluorescence micro-analysis (μ-SRXRF) and secondary ion mass spectrometry (SIMS) are compared for analyzing diffusion profiles of trace elements in two hydrous rhyolitic glasses (1.87 and 5.00wt% H2O). In order to verify the results, laser ablation coupled to inductively coupled plasma optical emission (LA-ICP-OES) has been used on one sample. Samples were produced by diffusion couple experiments performed in an internally heated gas pressure vessel at 1200°C and 500 MPa. One half of each couple was doped with 24 trace elements representing different geochemical groups: low field strength elements (Rb, Sr, Ba), transition metals (Cr, Co, Ni, Cu, Zn), rare earth elements (La, Ce, Nd, Sm, Eu, Gd, Er, Yb) + Y, high field strength elements (V, Zr, Nb, Hf, Ta) and main group elements (Ge, Sn). Several profiles were measured with both p-SRXRF and SIMS on both samples. In principle, concentrations of all elements can be extracted simultaneously from a single SRXRF spectrum. However, some trace elements could not be reliably quantified with our analytical system: Ta and Pb (used for detector collimator material), Ti, V (low energy of Kα, Co (Kα-peak overlapping with Fe K β-peak) and Cr, Ni, Cu, Zn (overlapping with 1-lines of REEs). In contrast, SIMS analyses measure each element sequentially. Hence, not all elements of the large total set of trace elements could be analyzed in a single run. Some elements requiring a high mass resolution (NaSi interfering with V, CaO interfering with Ni) or having low yields (Sn) were not profiled. Multi ple diffusivities derived from V-SRXRF and SIMS profiles are in very good agreement for most elements. In general, the trace element diffusivity decreases with increasing valence state, e.g. in sample D22 containing 1.87 wt% H 2O from log D = -10.80 for the monovalent Rb to log D=-13.34 for the tetravalent Zr (Din m2/s). By increasing the water content in sample D18 to 5.00 wt%, diffusion coefficients increase approximately by one order of magnitude for all elements studied.

KW - Rhyolite

KW - SIMS

KW - Synchroton radiation X-ray fluorescence microanalysis

KW - Trace element diffusion

UR - http://www.scopus.com/inward/record.url?scp=18444381750&partnerID=8YFLogxK

U2 - 10.1127/0935-1221/2005/0017-0233

DO - 10.1127/0935-1221/2005/0017-0233

M3 - Article

AN - SCOPUS:18444381750

VL - 17

SP - 233

EP - 242

JO - European journal of mineralogy

JF - European journal of mineralogy

SN - 0935-1221

IS - 2

ER -