Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Autoren

  • Dennis Kundrat
  • Andreas Schoob
  • Thomas Piskon
  • René Grässlin
  • Patrick J. Schuler
  • Thomas K. Hoffmann
  • Luder A. Kahrs
  • Tobias Ortmaier

Organisationseinheiten

Externe Organisationen

  • Universität Ulm
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Aufsatznummer8777169
Seiten (von - bis)145-157
Seitenumfang13
FachzeitschriftIEEE Transactions on Medical Robotics and Bionics
Jahrgang1
Ausgabenummer3
PublikationsstatusVeröffentlicht - 26 Juli 2019

Abstract

In the last decade, focused laser radiation has become an alternative instrumentation for soft tissue surgery due to precise ablation with simultaneous coagulation. However, adjustment of the focal position with respect to the tissue surface is crucial for optimal energy input and minimised trauma. Advantageous laser delivery without tissue contact poses challenges and demands for assistance. This contribution introduces three novel concepts for user assistance during focal adaptation in noncontact laser surgery with robotic teleoperation. Visual, haptic, and visuo-haptic feedback modalities are designed based on real-time scene reconstruction and laser-to-camera registration. The performance of proposed methods was assessed in a phantom study with positioning tasks under different assistance conditions using a teleoperated extensible continuum robot. Focal position error and task execution time were selected as performance metrics. In total, 15 subjects conducted trials under four different conditions ( N=240 ). A mean error reduction from 1.2 mm for nonassisted to 0.25 mm for visuo-haptic feedback was revealed and error outliers were eliminated efficiently. Mean execution times of below 13 s were achieved for applied conditions. Presented work decreased the error to one-fifth in comparison to absence of assistance in teleoperation. The results strongly motivate prospective integration to surgical devices for noncontact laser delivery.

ASJC Scopus Sachgebiete

Zitieren

Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery. / Kundrat, Dennis; Schoob, Andreas; Piskon, Thomas et al.
in: IEEE Transactions on Medical Robotics and Bionics, Jahrgang 1, Nr. 3, 8777169, 26.07.2019, S. 145-157.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Kundrat, D, Schoob, A, Piskon, T, Grässlin, R, Schuler, PJ, Hoffmann, TK, Kahrs, LA & Ortmaier, T 2019, 'Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery', IEEE Transactions on Medical Robotics and Bionics, Jg. 1, Nr. 3, 8777169, S. 145-157. https://doi.org/10.1109/tmrb.2019.2931438
Kundrat, D., Schoob, A., Piskon, T., Grässlin, R., Schuler, P. J., Hoffmann, T. K., Kahrs, L. A., & Ortmaier, T. (2019). Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery. IEEE Transactions on Medical Robotics and Bionics, 1(3), 145-157. Artikel 8777169. https://doi.org/10.1109/tmrb.2019.2931438
Kundrat D, Schoob A, Piskon T, Grässlin R, Schuler PJ, Hoffmann TK et al. Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery. IEEE Transactions on Medical Robotics and Bionics. 2019 Jul 26;1(3):145-157. 8777169. doi: 10.1109/tmrb.2019.2931438
Kundrat, Dennis ; Schoob, Andreas ; Piskon, Thomas et al. / Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery. in: IEEE Transactions on Medical Robotics and Bionics. 2019 ; Jahrgang 1, Nr. 3. S. 145-157.
Download
@article{b670e645f7d1418fbd786f7cd630df49,
title = "Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery",
abstract = "In the last decade, focused laser radiation has become an alternative instrumentation for soft tissue surgery due to precise ablation with simultaneous coagulation. However, adjustment of the focal position with respect to the tissue surface is crucial for optimal energy input and minimised trauma. Advantageous laser delivery without tissue contact poses challenges and demands for assistance. This contribution introduces three novel concepts for user assistance during focal adaptation in noncontact laser surgery with robotic teleoperation. Visual, haptic, and visuo-haptic feedback modalities are designed based on real-time scene reconstruction and laser-to-camera registration. The performance of proposed methods was assessed in a phantom study with positioning tasks under different assistance conditions using a teleoperated extensible continuum robot. Focal position error and task execution time were selected as performance metrics. In total, 15 subjects conducted trials under four different conditions ( N=240 ). A mean error reduction from 1.2 mm for nonassisted to 0.25 mm for visuo-haptic feedback was revealed and error outliers were eliminated efficiently. Mean execution times of below 13 s were achieved for applied conditions. Presented work decreased the error to one-fifth in comparison to absence of assistance in teleoperation. The results strongly motivate prospective integration to surgical devices for noncontact laser delivery.",
keywords = "Laser ablation, augmented reality, continuum robot, haptic cues, user assistance, user study, visual feedback",
author = "Dennis Kundrat and Andreas Schoob and Thomas Piskon and Ren{\'e} Gr{\"a}sslin and Schuler, {Patrick J.} and Hoffmann, {Thomas K.} and Kahrs, {Luder A.} and Tobias Ortmaier",
year = "2019",
month = jul,
day = "26",
doi = "10.1109/tmrb.2019.2931438",
language = "English",
volume = "1",
pages = "145--157",
number = "3",

}

Download

TY - JOUR

T1 - Toward Assistive Technologies for Focus Adjustment in Teleoperated Robotic Non-Contact Laser Surgery

AU - Kundrat, Dennis

AU - Schoob, Andreas

AU - Piskon, Thomas

AU - Grässlin, René

AU - Schuler, Patrick J.

AU - Hoffmann, Thomas K.

AU - Kahrs, Luder A.

AU - Ortmaier, Tobias

PY - 2019/7/26

Y1 - 2019/7/26

N2 - In the last decade, focused laser radiation has become an alternative instrumentation for soft tissue surgery due to precise ablation with simultaneous coagulation. However, adjustment of the focal position with respect to the tissue surface is crucial for optimal energy input and minimised trauma. Advantageous laser delivery without tissue contact poses challenges and demands for assistance. This contribution introduces three novel concepts for user assistance during focal adaptation in noncontact laser surgery with robotic teleoperation. Visual, haptic, and visuo-haptic feedback modalities are designed based on real-time scene reconstruction and laser-to-camera registration. The performance of proposed methods was assessed in a phantom study with positioning tasks under different assistance conditions using a teleoperated extensible continuum robot. Focal position error and task execution time were selected as performance metrics. In total, 15 subjects conducted trials under four different conditions ( N=240 ). A mean error reduction from 1.2 mm for nonassisted to 0.25 mm for visuo-haptic feedback was revealed and error outliers were eliminated efficiently. Mean execution times of below 13 s were achieved for applied conditions. Presented work decreased the error to one-fifth in comparison to absence of assistance in teleoperation. The results strongly motivate prospective integration to surgical devices for noncontact laser delivery.

AB - In the last decade, focused laser radiation has become an alternative instrumentation for soft tissue surgery due to precise ablation with simultaneous coagulation. However, adjustment of the focal position with respect to the tissue surface is crucial for optimal energy input and minimised trauma. Advantageous laser delivery without tissue contact poses challenges and demands for assistance. This contribution introduces three novel concepts for user assistance during focal adaptation in noncontact laser surgery with robotic teleoperation. Visual, haptic, and visuo-haptic feedback modalities are designed based on real-time scene reconstruction and laser-to-camera registration. The performance of proposed methods was assessed in a phantom study with positioning tasks under different assistance conditions using a teleoperated extensible continuum robot. Focal position error and task execution time were selected as performance metrics. In total, 15 subjects conducted trials under four different conditions ( N=240 ). A mean error reduction from 1.2 mm for nonassisted to 0.25 mm for visuo-haptic feedback was revealed and error outliers were eliminated efficiently. Mean execution times of below 13 s were achieved for applied conditions. Presented work decreased the error to one-fifth in comparison to absence of assistance in teleoperation. The results strongly motivate prospective integration to surgical devices for noncontact laser delivery.

KW - Laser ablation

KW - augmented reality

KW - continuum robot

KW - haptic cues

KW - user assistance

KW - user study

KW - visual feedback

UR - http://www.scopus.com/inward/record.url?scp=85079823214&partnerID=8YFLogxK

U2 - 10.1109/tmrb.2019.2931438

DO - 10.1109/tmrb.2019.2931438

M3 - Article

VL - 1

SP - 145

EP - 157

JO - IEEE Transactions on Medical Robotics and Bionics

JF - IEEE Transactions on Medical Robotics and Bionics

SN - 2576-3202

IS - 3

M1 - 8777169

ER -