The role of pressure and friction forces in automated insertion of cochlear implants

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Autoren

  • Max Fröhlich
  • Jaro Deutz
  • Matthias Wangenheim
  • Thomas S. Rau
  • Thomas Lenarz
  • Andrej Kral
  • Daniel Schurzig

Externe Organisationen

  • MED-EL Medical Electronics
  • Medizinische Hochschule Hannover (MHH)
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Aufsatznummer1430694
FachzeitschriftFrontiers in neurology
Jahrgang15
PublikationsstatusVeröffentlicht - 6 Aug. 2024

Abstract

Objectives: Despite the success of cochlear implant (CI) surgery for hearing restoration, reducing CI electrode insertion forces is an ongoing challenge with the goal to further reduce post-implantation hearing loss. While research in this field shows that both friction and quasistatic pressure forces occur during CI insertion, there is a lack of studies distinguishing between these origins. The present study was conducted to analyze the contribution of both force phenomena during automated CI insertion. Methods: Five MED-EL FLEX28 CI electrode arrays were inserted into both a regular and uncoiled version of the same average scala tympani (ST). Both ST models had a pressure release hole at the apical end, which was kept open or closed to quantify pressure forces. ST models were filled with different sodium dodecyl sulfate (SDS) lubricants (1, 5, and 10% SDS, water). The viscosity of lubricants was determined using a rheometer. Insertions were conducted with velocities ranging from v= 0.125 mm/s to 2.0 mm/s. Results: Viscosity of SDS lubricants at 20°C was 1.28, 1.96, and 2.51 mPas for 1, 5, and 10% SDS, respectively, which lies within the values reported for human perilymph. In the uncoiled ST model, forces remained within the noise floor (maximum: 0.049 × 10−3 N ± 1.5 × 10−3 N), indicating minimal contribution from quasistatic pressure. Conversely, forces using the regular, coiled ST model were at least an order of magnitude larger (minimum: Fmax = 28.95 × 10−3 N, v = 1 mm/s, 10% SDS), confirming that friction forces are the main contributor to total insertion forces. An N-way ANOVA revealed that both lubricant viscosity and insertion speed significantly reduce insertion forces (p < 0.001). Conclusion: For the first time, this study demonstrates that at realistic perilymph viscosities, quasistatic pressure forces minimally affect the total insertion force profile during insertion. Mixed friction is the main determinant, and significantly decreases with increaseing insertion speeds. This suggests that in clinical settings with similar ST geometries and surgical preparation, quasistatic pressure plays a subordinate role. Moreover, the findings indicate that managing the hydrodynamics of the cochlear environment, possibly through pre-surgical preparation or the use of specific lubricants, could effectively reduce insertion forces.

ASJC Scopus Sachgebiete

Zitieren

The role of pressure and friction forces in automated insertion of cochlear implants. / Fröhlich, Max; Deutz, Jaro; Wangenheim, Matthias et al.
in: Frontiers in neurology, Jahrgang 15, 1430694, 06.08.2024.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Fröhlich, M, Deutz, J, Wangenheim, M, Rau, TS, Lenarz, T, Kral, A & Schurzig, D 2024, 'The role of pressure and friction forces in automated insertion of cochlear implants', Frontiers in neurology, Jg. 15, 1430694. https://doi.org/10.3389/fneur.2024.1430694
Fröhlich, M., Deutz, J., Wangenheim, M., Rau, T. S., Lenarz, T., Kral, A., & Schurzig, D. (2024). The role of pressure and friction forces in automated insertion of cochlear implants. Frontiers in neurology, 15, Artikel 1430694. https://doi.org/10.3389/fneur.2024.1430694
Fröhlich M, Deutz J, Wangenheim M, Rau TS, Lenarz T, Kral A et al. The role of pressure and friction forces in automated insertion of cochlear implants. Frontiers in neurology. 2024 Aug 6;15:1430694. doi: 10.3389/fneur.2024.1430694
Fröhlich, Max ; Deutz, Jaro ; Wangenheim, Matthias et al. / The role of pressure and friction forces in automated insertion of cochlear implants. in: Frontiers in neurology. 2024 ; Jahrgang 15.
Download
@article{a06a63e772a84763906c15da2039f65f,
title = "The role of pressure and friction forces in automated insertion of cochlear implants",
abstract = "Objectives: Despite the success of cochlear implant (CI) surgery for hearing restoration, reducing CI electrode insertion forces is an ongoing challenge with the goal to further reduce post-implantation hearing loss. While research in this field shows that both friction and quasistatic pressure forces occur during CI insertion, there is a lack of studies distinguishing between these origins. The present study was conducted to analyze the contribution of both force phenomena during automated CI insertion. Methods: Five MED-EL FLEX28 CI electrode arrays were inserted into both a regular and uncoiled version of the same average scala tympani (ST). Both ST models had a pressure release hole at the apical end, which was kept open or closed to quantify pressure forces. ST models were filled with different sodium dodecyl sulfate (SDS) lubricants (1, 5, and 10% SDS, water). The viscosity of lubricants was determined using a rheometer. Insertions were conducted with velocities ranging from v= 0.125 mm/s to 2.0 mm/s. Results: Viscosity of SDS lubricants at 20°C was 1.28, 1.96, and 2.51 mPas for 1, 5, and 10% SDS, respectively, which lies within the values reported for human perilymph. In the uncoiled ST model, forces remained within the noise floor (maximum: 0.049 × 10−3 N ± 1.5 × 10−3 N), indicating minimal contribution from quasistatic pressure. Conversely, forces using the regular, coiled ST model were at least an order of magnitude larger (minimum: Fmax = 28.95 × 10−3 N, v = 1 mm/s, 10% SDS), confirming that friction forces are the main contributor to total insertion forces. An N-way ANOVA revealed that both lubricant viscosity and insertion speed significantly reduce insertion forces (p < 0.001). Conclusion: For the first time, this study demonstrates that at realistic perilymph viscosities, quasistatic pressure forces minimally affect the total insertion force profile during insertion. Mixed friction is the main determinant, and significantly decreases with increaseing insertion speeds. This suggests that in clinical settings with similar ST geometries and surgical preparation, quasistatic pressure plays a subordinate role. Moreover, the findings indicate that managing the hydrodynamics of the cochlear environment, possibly through pre-surgical preparation or the use of specific lubricants, could effectively reduce insertion forces.",
keywords = "cochlear implant, electrode array, friction, insertion force, insertion speed, perilymph viscosity, pressure, robotic surgery",
author = "Max Fr{\"o}hlich and Jaro Deutz and Matthias Wangenheim and Rau, {Thomas S.} and Thomas Lenarz and Andrej Kral and Daniel Schurzig",
note = "Publisher Copyright: Copyright {\textcopyright} 2024 Fr{\"o}hlich, Deutz, Wangenheim, Rau, Lenarz, Kral and Schurzig.",
year = "2024",
month = aug,
day = "6",
doi = "10.3389/fneur.2024.1430694",
language = "English",
volume = "15",
journal = "Frontiers in neurology",
issn = "1664-2295",
publisher = "Frontiers Media S.A.",

}

Download

TY - JOUR

T1 - The role of pressure and friction forces in automated insertion of cochlear implants

AU - Fröhlich, Max

AU - Deutz, Jaro

AU - Wangenheim, Matthias

AU - Rau, Thomas S.

AU - Lenarz, Thomas

AU - Kral, Andrej

AU - Schurzig, Daniel

N1 - Publisher Copyright: Copyright © 2024 Fröhlich, Deutz, Wangenheim, Rau, Lenarz, Kral and Schurzig.

PY - 2024/8/6

Y1 - 2024/8/6

N2 - Objectives: Despite the success of cochlear implant (CI) surgery for hearing restoration, reducing CI electrode insertion forces is an ongoing challenge with the goal to further reduce post-implantation hearing loss. While research in this field shows that both friction and quasistatic pressure forces occur during CI insertion, there is a lack of studies distinguishing between these origins. The present study was conducted to analyze the contribution of both force phenomena during automated CI insertion. Methods: Five MED-EL FLEX28 CI electrode arrays were inserted into both a regular and uncoiled version of the same average scala tympani (ST). Both ST models had a pressure release hole at the apical end, which was kept open or closed to quantify pressure forces. ST models were filled with different sodium dodecyl sulfate (SDS) lubricants (1, 5, and 10% SDS, water). The viscosity of lubricants was determined using a rheometer. Insertions were conducted with velocities ranging from v= 0.125 mm/s to 2.0 mm/s. Results: Viscosity of SDS lubricants at 20°C was 1.28, 1.96, and 2.51 mPas for 1, 5, and 10% SDS, respectively, which lies within the values reported for human perilymph. In the uncoiled ST model, forces remained within the noise floor (maximum: 0.049 × 10−3 N ± 1.5 × 10−3 N), indicating minimal contribution from quasistatic pressure. Conversely, forces using the regular, coiled ST model were at least an order of magnitude larger (minimum: Fmax = 28.95 × 10−3 N, v = 1 mm/s, 10% SDS), confirming that friction forces are the main contributor to total insertion forces. An N-way ANOVA revealed that both lubricant viscosity and insertion speed significantly reduce insertion forces (p < 0.001). Conclusion: For the first time, this study demonstrates that at realistic perilymph viscosities, quasistatic pressure forces minimally affect the total insertion force profile during insertion. Mixed friction is the main determinant, and significantly decreases with increaseing insertion speeds. This suggests that in clinical settings with similar ST geometries and surgical preparation, quasistatic pressure plays a subordinate role. Moreover, the findings indicate that managing the hydrodynamics of the cochlear environment, possibly through pre-surgical preparation or the use of specific lubricants, could effectively reduce insertion forces.

AB - Objectives: Despite the success of cochlear implant (CI) surgery for hearing restoration, reducing CI electrode insertion forces is an ongoing challenge with the goal to further reduce post-implantation hearing loss. While research in this field shows that both friction and quasistatic pressure forces occur during CI insertion, there is a lack of studies distinguishing between these origins. The present study was conducted to analyze the contribution of both force phenomena during automated CI insertion. Methods: Five MED-EL FLEX28 CI electrode arrays were inserted into both a regular and uncoiled version of the same average scala tympani (ST). Both ST models had a pressure release hole at the apical end, which was kept open or closed to quantify pressure forces. ST models were filled with different sodium dodecyl sulfate (SDS) lubricants (1, 5, and 10% SDS, water). The viscosity of lubricants was determined using a rheometer. Insertions were conducted with velocities ranging from v= 0.125 mm/s to 2.0 mm/s. Results: Viscosity of SDS lubricants at 20°C was 1.28, 1.96, and 2.51 mPas for 1, 5, and 10% SDS, respectively, which lies within the values reported for human perilymph. In the uncoiled ST model, forces remained within the noise floor (maximum: 0.049 × 10−3 N ± 1.5 × 10−3 N), indicating minimal contribution from quasistatic pressure. Conversely, forces using the regular, coiled ST model were at least an order of magnitude larger (minimum: Fmax = 28.95 × 10−3 N, v = 1 mm/s, 10% SDS), confirming that friction forces are the main contributor to total insertion forces. An N-way ANOVA revealed that both lubricant viscosity and insertion speed significantly reduce insertion forces (p < 0.001). Conclusion: For the first time, this study demonstrates that at realistic perilymph viscosities, quasistatic pressure forces minimally affect the total insertion force profile during insertion. Mixed friction is the main determinant, and significantly decreases with increaseing insertion speeds. This suggests that in clinical settings with similar ST geometries and surgical preparation, quasistatic pressure plays a subordinate role. Moreover, the findings indicate that managing the hydrodynamics of the cochlear environment, possibly through pre-surgical preparation or the use of specific lubricants, could effectively reduce insertion forces.

KW - cochlear implant

KW - electrode array

KW - friction

KW - insertion force

KW - insertion speed

KW - perilymph viscosity

KW - pressure

KW - robotic surgery

UR - http://www.scopus.com/inward/record.url?scp=85201572952&partnerID=8YFLogxK

U2 - 10.3389/fneur.2024.1430694

DO - 10.3389/fneur.2024.1430694

M3 - Article

AN - SCOPUS:85201572952

VL - 15

JO - Frontiers in neurology

JF - Frontiers in neurology

SN - 1664-2295

M1 - 1430694

ER -