Temperature control with internally applied cooling in solid material drilling: an experimental, biomechanical study

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Autorschaft

  • Stephan Brand
  • Johannes Klotz
  • Maximilian Petri
  • Max Ettinger
  • Thomas Hassel
  • Christian Krettek
  • Thomas Goesling
  • Friedrich Wilhelm Bach

Organisationseinheiten

Externe Organisationen

  • Medizinische Hochschule Hannover (MHH)
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Seiten (von - bis)1355-1361
Seitenumfang7
FachzeitschriftInternational orthopaedics
Jahrgang37
Ausgabenummer7
Frühes Online-Datum20 März 2013
PublikationsstatusVeröffentlicht - Juli 2013

Abstract

Purpose: The purpose of this study was to evaluate the different temperature levels while drilling solid materials and to compare different cooling solutions for possible temperature control. An additional purpose was to develop an internal cooling device which can be connected to routinely used manual drilling devices in trauma surgery. Methods: Drilling was performed on a straight hip stem implanted in bovine femora without cooling, with externally applied cooling and with a newly developed internal cooling device. Temperature changes were measured by seven thermocouples arranged near the borehole. Additionally, thermographic scans were performed during drilling. Results: Drilling without cooling leads to an immediate increase in temperature to levels of thermal osteonecrosis (over 200 C). With externally applied cooling temperatures were decreased, but were still up to a tissue damaging 85 C. Internally applied cooling led to a temperature decrease to tissue-preserving levels during the drilling procedure (24.7 C). Conclusion: Internal cooling with HPC-drillers lowered the measured temperatures to non-tissue damaging temperatures and should avoid structural tissue damage.

ASJC Scopus Sachgebiete

Zitieren

Temperature control with internally applied cooling in solid material drilling: an experimental, biomechanical study. / Brand, Stephan; Klotz, Johannes; Petri, Maximilian et al.
in: International orthopaedics, Jahrgang 37, Nr. 7, 07.2013, S. 1355-1361.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Brand, S, Klotz, J, Petri, M, Ettinger, M, Hassel, T, Krettek, C, Goesling, T & Bach, FW 2013, 'Temperature control with internally applied cooling in solid material drilling: an experimental, biomechanical study', International orthopaedics, Jg. 37, Nr. 7, S. 1355-1361. https://doi.org/10.1007/s00264-013-1850-4
Brand, S., Klotz, J., Petri, M., Ettinger, M., Hassel, T., Krettek, C., Goesling, T., & Bach, F. W. (2013). Temperature control with internally applied cooling in solid material drilling: an experimental, biomechanical study. International orthopaedics, 37(7), 1355-1361. https://doi.org/10.1007/s00264-013-1850-4
Brand S, Klotz J, Petri M, Ettinger M, Hassel T, Krettek C et al. Temperature control with internally applied cooling in solid material drilling: an experimental, biomechanical study. International orthopaedics. 2013 Jul;37(7):1355-1361. Epub 2013 Mär 20. doi: 10.1007/s00264-013-1850-4
Brand, Stephan ; Klotz, Johannes ; Petri, Maximilian et al. / Temperature control with internally applied cooling in solid material drilling : an experimental, biomechanical study. in: International orthopaedics. 2013 ; Jahrgang 37, Nr. 7. S. 1355-1361.
Download
@article{e4a07f1072af4c33afa2fe76f92d3382,
title = "Temperature control with internally applied cooling in solid material drilling: an experimental, biomechanical study",
abstract = "Purpose: The purpose of this study was to evaluate the different temperature levels while drilling solid materials and to compare different cooling solutions for possible temperature control. An additional purpose was to develop an internal cooling device which can be connected to routinely used manual drilling devices in trauma surgery. Methods: Drilling was performed on a straight hip stem implanted in bovine femora without cooling, with externally applied cooling and with a newly developed internal cooling device. Temperature changes were measured by seven thermocouples arranged near the borehole. Additionally, thermographic scans were performed during drilling. Results: Drilling without cooling leads to an immediate increase in temperature to levels of thermal osteonecrosis (over 200 C). With externally applied cooling temperatures were decreased, but were still up to a tissue damaging 85 C. Internally applied cooling led to a temperature decrease to tissue-preserving levels during the drilling procedure (24.7 C). Conclusion: Internal cooling with HPC-drillers lowered the measured temperatures to non-tissue damaging temperatures and should avoid structural tissue damage.",
author = "Stephan Brand and Johannes Klotz and Maximilian Petri and Max Ettinger and Thomas Hassel and Christian Krettek and Thomas Goesling and Bach, {Friedrich Wilhelm}",
year = "2013",
month = jul,
doi = "10.1007/s00264-013-1850-4",
language = "English",
volume = "37",
pages = "1355--1361",
journal = "International orthopaedics",
issn = "0341-2695",
publisher = "Springer Verlag",
number = "7",

}

Download

TY - JOUR

T1 - Temperature control with internally applied cooling in solid material drilling

T2 - an experimental, biomechanical study

AU - Brand, Stephan

AU - Klotz, Johannes

AU - Petri, Maximilian

AU - Ettinger, Max

AU - Hassel, Thomas

AU - Krettek, Christian

AU - Goesling, Thomas

AU - Bach, Friedrich Wilhelm

PY - 2013/7

Y1 - 2013/7

N2 - Purpose: The purpose of this study was to evaluate the different temperature levels while drilling solid materials and to compare different cooling solutions for possible temperature control. An additional purpose was to develop an internal cooling device which can be connected to routinely used manual drilling devices in trauma surgery. Methods: Drilling was performed on a straight hip stem implanted in bovine femora without cooling, with externally applied cooling and with a newly developed internal cooling device. Temperature changes were measured by seven thermocouples arranged near the borehole. Additionally, thermographic scans were performed during drilling. Results: Drilling without cooling leads to an immediate increase in temperature to levels of thermal osteonecrosis (over 200 C). With externally applied cooling temperatures were decreased, but were still up to a tissue damaging 85 C. Internally applied cooling led to a temperature decrease to tissue-preserving levels during the drilling procedure (24.7 C). Conclusion: Internal cooling with HPC-drillers lowered the measured temperatures to non-tissue damaging temperatures and should avoid structural tissue damage.

AB - Purpose: The purpose of this study was to evaluate the different temperature levels while drilling solid materials and to compare different cooling solutions for possible temperature control. An additional purpose was to develop an internal cooling device which can be connected to routinely used manual drilling devices in trauma surgery. Methods: Drilling was performed on a straight hip stem implanted in bovine femora without cooling, with externally applied cooling and with a newly developed internal cooling device. Temperature changes were measured by seven thermocouples arranged near the borehole. Additionally, thermographic scans were performed during drilling. Results: Drilling without cooling leads to an immediate increase in temperature to levels of thermal osteonecrosis (over 200 C). With externally applied cooling temperatures were decreased, but were still up to a tissue damaging 85 C. Internally applied cooling led to a temperature decrease to tissue-preserving levels during the drilling procedure (24.7 C). Conclusion: Internal cooling with HPC-drillers lowered the measured temperatures to non-tissue damaging temperatures and should avoid structural tissue damage.

UR - http://www.scopus.com/inward/record.url?scp=84879410803&partnerID=8YFLogxK

U2 - 10.1007/s00264-013-1850-4

DO - 10.1007/s00264-013-1850-4

M3 - Article

C2 - 23512602

AN - SCOPUS:84879410803

VL - 37

SP - 1355

EP - 1361

JO - International orthopaedics

JF - International orthopaedics

SN - 0341-2695

IS - 7

ER -