Abstract
We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery.
ASJC Scopus Sachgebiete
- Werkstoffwissenschaften (insg.)
- Elektronische, optische und magnetische Materialien
- Physik und Astronomie (insg.)
- Atom- und Molekularphysik sowie Optik
Zitieren
- Standard
- Harvard
- Apa
- Vancouver
- BibTex
- RIS
in: Journal of Optics (United Kingdom), Jahrgang 18, Nr. 1, 015301, 01.12.2015.
Publikation: Beitrag in Fachzeitschrift › Artikel › Forschung › Peer-Review
}
TY - JOUR
T1 - One step geometrical calibration method for optical coherence tomography
AU - Díaz, Jesús Díaz
AU - Stritzel, Jenny
AU - Rahlves, Maik
AU - Majdani, Omid
AU - Reithmeier, Eduard
AU - Ortmaier, Tobias
AU - Roth, Bernhard
N1 - Publisher Copyright: © 2016 IOP Publishing Ltd. Copyright: Copyright 2020 Elsevier B.V., All rights reserved.
PY - 2015/12/1
Y1 - 2015/12/1
N2 - We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery.
AB - We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery.
KW - calibration standards
KW - geometrical calibration
KW - optical coherence tomography
KW - threedimensional measurement
UR - http://www.scopus.com/inward/record.url?scp=84951742126&partnerID=8YFLogxK
U2 - 10.1088/2040-8978/18/1/015301
DO - 10.1088/2040-8978/18/1/015301
M3 - Article
AN - SCOPUS:84951742126
VL - 18
JO - Journal of Optics (United Kingdom)
JF - Journal of Optics (United Kingdom)
SN - 2040-8978
IS - 1
M1 - 015301
ER -