Imperfect refractive index matching in scanning laser optical tomography and a method for digital correction

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Autoren

  • Ole Hill
  • Merve Wollweber
  • Tobias Biermann
  • Tammo Ripken
  • Roland Lachmayer

Organisationseinheiten

Externe Organisationen

  • Laser Zentrum Hannover e.V. (LZH)
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Seitenumfang15
FachzeitschriftJournal of biomedical optics
Jahrgang29
Ausgabenummer6
PublikationsstatusVeröffentlicht - 15 Mai 2024

Abstract

Significance: Scanning laser optical tomography (SLOT) is a volumetric multi-modal imaging technique that is comparable to optical projection tomography and computer tomography. Image quality is crucially dependent on matching the refractive indexes (RIs) of the sample and surrounding medium, but RI matching often requires some effort and is never perfect. Aim: Reducing the burden of RI matching between the immersion medium and sample in biomedical imaging is a challenging and interesting task. We aim at implementing a post processing strategy for correcting SLOT measurements that have errors caused by RI mismatch. Approach: To better understand the problems with poorly matched Ris, simulated SLOT measurements with imperfect RI matching of the sample and medium are performed and presented here. A method to correct distorted measurements was developed and is presented and evaluated. This method is then applied to a sample containing fluorescent polystyrene beads and a sample made of olydimethylsiloxane with embedded fluorescent nanoparticles. Results: From the simulations, it is evident that measurements with an RI mismatch larger than 0.02 and no correction yield considerably worse results compared to perfectly matched measurements. RI mismatches larger than 0.05 make it almost impossible to resolve finer details and structures. By contrast, the simulations imply that a measurement with an RI mismatch of up to 0.1 can still yield reasonable results if the presented correction method is applied. The experiments validate the simulated results for an RI mismatch of about 0.09. Conclusions: The method significantly improves the SLOT image quality for samples with imperfectly matched Ris. Although the absolutely best imaging quality will be achieved with perfect RI matching, these results pave the way for imaging in SLOT with RI mismatches while maintaining high image quality.

ASJC Scopus Sachgebiete

Zitieren

Imperfect refractive index matching in scanning laser optical tomography and a method for digital correction. / Hill, Ole; Wollweber, Merve; Biermann, Tobias et al.
in: Journal of biomedical optics, Jahrgang 29, Nr. 6, 15.05.2024.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Hill O, Wollweber M, Biermann T, Ripken T, Lachmayer R. Imperfect refractive index matching in scanning laser optical tomography and a method for digital correction. Journal of biomedical optics. 2024 Mai 15;29(6). doi: 10.1117/1.JBO.29.6.066004
Hill, Ole ; Wollweber, Merve ; Biermann, Tobias et al. / Imperfect refractive index matching in scanning laser optical tomography and a method for digital correction. in: Journal of biomedical optics. 2024 ; Jahrgang 29, Nr. 6.
Download
@article{190430b75b974798be5f20304a3e5633,
title = "Imperfect refractive index matching in scanning laser optical tomography and a method for digital correction",
abstract = "Significance: Scanning laser optical tomography (SLOT) is a volumetric multi-modal imaging technique that is comparable to optical projection tomography and computer tomography. Image quality is crucially dependent on matching the refractive indexes (RIs) of the sample and surrounding medium, but RI matching often requires some effort and is never perfect. Aim: Reducing the burden of RI matching between the immersion medium and sample in biomedical imaging is a challenging and interesting task. We aim at implementing a post processing strategy for correcting SLOT measurements that have errors caused by RI mismatch. Approach: To better understand the problems with poorly matched Ris, simulated SLOT measurements with imperfect RI matching of the sample and medium are performed and presented here. A method to correct distorted measurements was developed and is presented and evaluated. This method is then applied to a sample containing fluorescent polystyrene beads and a sample made of olydimethylsiloxane with embedded fluorescent nanoparticles. Results: From the simulations, it is evident that measurements with an RI mismatch larger than 0.02 and no correction yield considerably worse results compared to perfectly matched measurements. RI mismatches larger than 0.05 make it almost impossible to resolve finer details and structures. By contrast, the simulations imply that a measurement with an RI mismatch of up to 0.1 can still yield reasonable results if the presented correction method is applied. The experiments validate the simulated results for an RI mismatch of about 0.09. Conclusions: The method significantly improves the SLOT image quality for samples with imperfectly matched Ris. Although the absolutely best imaging quality will be achieved with perfect RI matching, these results pave the way for imaging in SLOT with RI mismatches while maintaining high image quality.",
keywords = "digital microscopy, non-destructive, scanning laser optical tomography, tomography",
author = "Ole Hill and Merve Wollweber and Tobias Biermann and Tammo Ripken and Roland Lachmayer",
note = "Publisher Copyright: {\textcopyright} 2024 The Authors.",
year = "2024",
month = may,
day = "15",
doi = "10.1117/1.JBO.29.6.066004",
language = "English",
volume = "29",
journal = "Journal of biomedical optics",
issn = "1083-3668",
publisher = "SPIE",
number = "6",

}

Download

TY - JOUR

T1 - Imperfect refractive index matching in scanning laser optical tomography and a method for digital correction

AU - Hill, Ole

AU - Wollweber, Merve

AU - Biermann, Tobias

AU - Ripken, Tammo

AU - Lachmayer, Roland

N1 - Publisher Copyright: © 2024 The Authors.

PY - 2024/5/15

Y1 - 2024/5/15

N2 - Significance: Scanning laser optical tomography (SLOT) is a volumetric multi-modal imaging technique that is comparable to optical projection tomography and computer tomography. Image quality is crucially dependent on matching the refractive indexes (RIs) of the sample and surrounding medium, but RI matching often requires some effort and is never perfect. Aim: Reducing the burden of RI matching between the immersion medium and sample in biomedical imaging is a challenging and interesting task. We aim at implementing a post processing strategy for correcting SLOT measurements that have errors caused by RI mismatch. Approach: To better understand the problems with poorly matched Ris, simulated SLOT measurements with imperfect RI matching of the sample and medium are performed and presented here. A method to correct distorted measurements was developed and is presented and evaluated. This method is then applied to a sample containing fluorescent polystyrene beads and a sample made of olydimethylsiloxane with embedded fluorescent nanoparticles. Results: From the simulations, it is evident that measurements with an RI mismatch larger than 0.02 and no correction yield considerably worse results compared to perfectly matched measurements. RI mismatches larger than 0.05 make it almost impossible to resolve finer details and structures. By contrast, the simulations imply that a measurement with an RI mismatch of up to 0.1 can still yield reasonable results if the presented correction method is applied. The experiments validate the simulated results for an RI mismatch of about 0.09. Conclusions: The method significantly improves the SLOT image quality for samples with imperfectly matched Ris. Although the absolutely best imaging quality will be achieved with perfect RI matching, these results pave the way for imaging in SLOT with RI mismatches while maintaining high image quality.

AB - Significance: Scanning laser optical tomography (SLOT) is a volumetric multi-modal imaging technique that is comparable to optical projection tomography and computer tomography. Image quality is crucially dependent on matching the refractive indexes (RIs) of the sample and surrounding medium, but RI matching often requires some effort and is never perfect. Aim: Reducing the burden of RI matching between the immersion medium and sample in biomedical imaging is a challenging and interesting task. We aim at implementing a post processing strategy for correcting SLOT measurements that have errors caused by RI mismatch. Approach: To better understand the problems with poorly matched Ris, simulated SLOT measurements with imperfect RI matching of the sample and medium are performed and presented here. A method to correct distorted measurements was developed and is presented and evaluated. This method is then applied to a sample containing fluorescent polystyrene beads and a sample made of olydimethylsiloxane with embedded fluorescent nanoparticles. Results: From the simulations, it is evident that measurements with an RI mismatch larger than 0.02 and no correction yield considerably worse results compared to perfectly matched measurements. RI mismatches larger than 0.05 make it almost impossible to resolve finer details and structures. By contrast, the simulations imply that a measurement with an RI mismatch of up to 0.1 can still yield reasonable results if the presented correction method is applied. The experiments validate the simulated results for an RI mismatch of about 0.09. Conclusions: The method significantly improves the SLOT image quality for samples with imperfectly matched Ris. Although the absolutely best imaging quality will be achieved with perfect RI matching, these results pave the way for imaging in SLOT with RI mismatches while maintaining high image quality.

KW - digital microscopy

KW - non-destructive

KW - scanning laser optical tomography

KW - tomography

UR - http://www.scopus.com/inward/record.url?scp=85193348137&partnerID=8YFLogxK

U2 - 10.1117/1.JBO.29.6.066004

DO - 10.1117/1.JBO.29.6.066004

M3 - Article

C2 - 38751827

AN - SCOPUS:85193348137

VL - 29

JO - Journal of biomedical optics

JF - Journal of biomedical optics

SN - 1083-3668

IS - 6

ER -