Abstract
In this paper, new on-line sensors based on pH-sensitive field-effect transitors (pH-FET) are presented for the measurement of ammonia and carbon dioxide during different cultivation processes. Both systems show a linear response between 2.7x10-4 and 2.7x10-2moll-1 for ammonia and 2.2x10-4 and 1.5x10-2moll-1 for dissolved carbon dioxide. A similar system can be used for the determination of enzyme activities. Copyright (C) 1998 Elsevier Science B.V.
ASJC Scopus Sachgebiete
- Chemie (insg.)
- Analytische Chemie
- Biochemie, Genetik und Molekularbiologie (insg.)
- Biochemie
- Umweltwissenschaften (insg.)
- Umweltchemie
- Chemie (insg.)
- Spektroskopie
Zitieren
- Standard
- Harvard
- Apa
- Vancouver
- BibTex
- RIS
in: Analytica chimica acta, Jahrgang 373, Nr. 2-3, 20.10.1998, S. 253-259.
Publikation: Beitrag in Fachzeitschrift › Artikel › Forschung › Peer-Review
}
TY - JOUR
T1 - Gas sensitive field-effect transistors: applications to the monitoring of biotechnological processes
AU - Ulber, Roland
AU - Kullick, Thomas
AU - Thordsen, Oliver
AU - Ritzka, Adnan
AU - Irtel, Florian
AU - Hitzmann, Bernd
AU - Scheper, Thomas
PY - 1998/10/20
Y1 - 1998/10/20
N2 - In this paper, new on-line sensors based on pH-sensitive field-effect transitors (pH-FET) are presented for the measurement of ammonia and carbon dioxide during different cultivation processes. Both systems show a linear response between 2.7x10-4 and 2.7x10-2moll-1 for ammonia and 2.2x10-4 and 1.5x10-2moll-1 for dissolved carbon dioxide. A similar system can be used for the determination of enzyme activities. Copyright (C) 1998 Elsevier Science B.V.
AB - In this paper, new on-line sensors based on pH-sensitive field-effect transitors (pH-FET) are presented for the measurement of ammonia and carbon dioxide during different cultivation processes. Both systems show a linear response between 2.7x10-4 and 2.7x10-2moll-1 for ammonia and 2.2x10-4 and 1.5x10-2moll-1 for dissolved carbon dioxide. A similar system can be used for the determination of enzyme activities. Copyright (C) 1998 Elsevier Science B.V.
KW - Ammonia
KW - Bioprocess monitoring and control
KW - Carbon dioxide
KW - Dissolved gas sensor
KW - Enzyme activity
KW - pH-FET
UR - http://www.scopus.com/inward/record.url?scp=0032508898&partnerID=8YFLogxK
U2 - 10.1016/S0003-2670(98)00424-3
DO - 10.1016/S0003-2670(98)00424-3
M3 - Article
AN - SCOPUS:0032508898
VL - 373
SP - 253
EP - 259
JO - Analytica chimica acta
JF - Analytica chimica acta
SN - 0003-2670
IS - 2-3
ER -