Functional quality assessment of whole-body vibration training devices based on instantaneous amplitude and frequency of photogrammetric vibration measurements

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Autoren

  • Sanam Moghaddamnia
  • Robin Rofallski
  • Thomas Luhmann
  • Tobias S. Kaeding

Organisationseinheiten

Externe Organisationen

  • Türkisch-Deutsche Universität
  • Jade Hochschule Wilhelmshaven/Oldenburg/Elsfleth
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Aufsatznummer103935
FachzeitschriftMedical Engineering and Physics
Jahrgang111
Frühes Online-Datum29 Nov. 2022
PublikationsstatusVeröffentlicht - Jan. 2023

Abstract

The practical use of whole-body vibration training (WBVT) and such research may be negatively influenced by generated vibrations with amplitudes, frequencies, and/or patterns that deviate from preset adjustments on WBVT devices. This study examined whether prolonged regular use can generate respective deviations. Four WBVT devices, used for 19 months in a research project on the effects of WBVT, were analyzed using photogrammetry before start of the research project and after 19 months. Divergences between preset and measured amplitudes and frequencies were calculated for all measurements. To quantify how well the output of devices correlates with the target setting, the vibration characteristics were calculated. In particular, exact long-term measurements related to the vibration amplitude is conducted and analyzed for the first time, which has been found as an important measure of the device functional quality. One device had a significantly (p<0.01) larger machine run time than the other three. This one showed the most pronounced signs of functional impairments concerning instantaneous amplitudes, frequencies and the mode of vibration after prolonged use. These results based on photometric measurements underline again that prolonged use can result in divergences between preset and actual applied amplitudes, frequencies, mode of vibration and other accuracy measurement metrics.

ASJC Scopus Sachgebiete

Zitieren

Functional quality assessment of whole-body vibration training devices based on instantaneous amplitude and frequency of photogrammetric vibration measurements. / Moghaddamnia, Sanam; Rofallski, Robin; Luhmann, Thomas et al.
in: Medical Engineering and Physics, Jahrgang 111, 103935, 01.2023.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Moghaddamnia S, Rofallski R, Luhmann T, Kaeding TS. Functional quality assessment of whole-body vibration training devices based on instantaneous amplitude and frequency of photogrammetric vibration measurements. Medical Engineering and Physics. 2023 Jan;111:103935. Epub 2022 Nov 29. doi: 10.1016/j.medengphy.2022.103935
Download
@article{29e1bb6d5ad94218bf0724bd90ace184,
title = "Functional quality assessment of whole-body vibration training devices based on instantaneous amplitude and frequency of photogrammetric vibration measurements",
abstract = "The practical use of whole-body vibration training (WBVT) and such research may be negatively influenced by generated vibrations with amplitudes, frequencies, and/or patterns that deviate from preset adjustments on WBVT devices. This study examined whether prolonged regular use can generate respective deviations. Four WBVT devices, used for 19 months in a research project on the effects of WBVT, were analyzed using photogrammetry before start of the research project and after 19 months. Divergences between preset and measured amplitudes and frequencies were calculated for all measurements. To quantify how well the output of devices correlates with the target setting, the vibration characteristics were calculated. In particular, exact long-term measurements related to the vibration amplitude is conducted and analyzed for the first time, which has been found as an important measure of the device functional quality. One device had a significantly (p<0.01) larger machine run time than the other three. This one showed the most pronounced signs of functional impairments concerning instantaneous amplitudes, frequencies and the mode of vibration after prolonged use. These results based on photometric measurements underline again that prolonged use can result in divergences between preset and actual applied amplitudes, frequencies, mode of vibration and other accuracy measurement metrics.",
keywords = "Functional quality, Photogrammetric measurements, Prolonged use, Vibration characteristics, Whole-body vibration training",
author = "Sanam Moghaddamnia and Robin Rofallski and Thomas Luhmann and Kaeding, {Tobias S.}",
year = "2023",
month = jan,
doi = "10.1016/j.medengphy.2022.103935",
language = "English",
volume = "111",
journal = "Medical Engineering and Physics",
issn = "1350-4533",
publisher = "Elsevier BV",

}

Download

TY - JOUR

T1 - Functional quality assessment of whole-body vibration training devices based on instantaneous amplitude and frequency of photogrammetric vibration measurements

AU - Moghaddamnia, Sanam

AU - Rofallski, Robin

AU - Luhmann, Thomas

AU - Kaeding, Tobias S.

PY - 2023/1

Y1 - 2023/1

N2 - The practical use of whole-body vibration training (WBVT) and such research may be negatively influenced by generated vibrations with amplitudes, frequencies, and/or patterns that deviate from preset adjustments on WBVT devices. This study examined whether prolonged regular use can generate respective deviations. Four WBVT devices, used for 19 months in a research project on the effects of WBVT, were analyzed using photogrammetry before start of the research project and after 19 months. Divergences between preset and measured amplitudes and frequencies were calculated for all measurements. To quantify how well the output of devices correlates with the target setting, the vibration characteristics were calculated. In particular, exact long-term measurements related to the vibration amplitude is conducted and analyzed for the first time, which has been found as an important measure of the device functional quality. One device had a significantly (p<0.01) larger machine run time than the other three. This one showed the most pronounced signs of functional impairments concerning instantaneous amplitudes, frequencies and the mode of vibration after prolonged use. These results based on photometric measurements underline again that prolonged use can result in divergences between preset and actual applied amplitudes, frequencies, mode of vibration and other accuracy measurement metrics.

AB - The practical use of whole-body vibration training (WBVT) and such research may be negatively influenced by generated vibrations with amplitudes, frequencies, and/or patterns that deviate from preset adjustments on WBVT devices. This study examined whether prolonged regular use can generate respective deviations. Four WBVT devices, used for 19 months in a research project on the effects of WBVT, were analyzed using photogrammetry before start of the research project and after 19 months. Divergences between preset and measured amplitudes and frequencies were calculated for all measurements. To quantify how well the output of devices correlates with the target setting, the vibration characteristics were calculated. In particular, exact long-term measurements related to the vibration amplitude is conducted and analyzed for the first time, which has been found as an important measure of the device functional quality. One device had a significantly (p<0.01) larger machine run time than the other three. This one showed the most pronounced signs of functional impairments concerning instantaneous amplitudes, frequencies and the mode of vibration after prolonged use. These results based on photometric measurements underline again that prolonged use can result in divergences between preset and actual applied amplitudes, frequencies, mode of vibration and other accuracy measurement metrics.

KW - Functional quality

KW - Photogrammetric measurements

KW - Prolonged use

KW - Vibration characteristics

KW - Whole-body vibration training

UR - http://www.scopus.com/inward/record.url?scp=85144083452&partnerID=8YFLogxK

U2 - 10.1016/j.medengphy.2022.103935

DO - 10.1016/j.medengphy.2022.103935

M3 - Article

AN - SCOPUS:85144083452

VL - 111

JO - Medical Engineering and Physics

JF - Medical Engineering and Physics

SN - 1350-4533

M1 - 103935

ER -