Evaluation of RF localization for deep brain implants

Publikation: Beitrag in Buch/Bericht/Sammelwerk/KonferenzbandAufsatz in KonferenzbandForschungPeer-Review

Autorschaft

  • D. Manteuffel
  • P. A. Hoeher
  • M. Mehdorn

Externe Organisationen

  • Christian-Albrechts-Universität zu Kiel (CAU)
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Titel des SammelwerksProceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011
Seiten2372-2376
Seitenumfang5
PublikationsstatusVeröffentlicht - 2011
Extern publiziertJa
Veranstaltung5th European Conference on Antennas and Propagation, EUCAP 2011 - Rome, Italien
Dauer: 10 Apr. 201115 Apr. 2011

Publikationsreihe

NameProceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011

Abstract

This paper presents first results of an advanced study based on numerical simulations concerning the applicability of RF techniques for the localization of functional medical probes located in the ventricular system of the brain. The functional implant contains a dipole antenna that transmits short pulses in the frequency range of 3.1 < f[GHz] < 5.1. The localization concept is based on angle of arrival determination at several reference stations located around the skull of the phantom. The investigation of three test cases shows that the accuracy is significantly impaired if the spherical wave front is overlapped by a guided wave in the bone layer of the skull. Reasonable localization accuracy can be achieved if the reference stations are located at positions where the skull bone is thin.

ASJC Scopus Sachgebiete

Zitieren

Evaluation of RF localization for deep brain implants. / Manteuffel, D.; Hoeher, P. A.; Mehdorn, M.
Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011. 2011. S. 2372-2376 5781910 (Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011).

Publikation: Beitrag in Buch/Bericht/Sammelwerk/KonferenzbandAufsatz in KonferenzbandForschungPeer-Review

Manteuffel, D, Hoeher, PA & Mehdorn, M 2011, Evaluation of RF localization for deep brain implants. in Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011., 5781910, Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011, S. 2372-2376, 5th European Conference on Antennas and Propagation, EUCAP 2011, Rome, Italien, 10 Apr. 2011.
Manteuffel, D., Hoeher, P. A., & Mehdorn, M. (2011). Evaluation of RF localization for deep brain implants. In Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011 (S. 2372-2376). Artikel 5781910 (Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011).
Manteuffel D, Hoeher PA, Mehdorn M. Evaluation of RF localization for deep brain implants. in Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011. 2011. S. 2372-2376. 5781910. (Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011).
Manteuffel, D. ; Hoeher, P. A. ; Mehdorn, M. / Evaluation of RF localization for deep brain implants. Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011. 2011. S. 2372-2376 (Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011).
Download
@inproceedings{96864fd75ac347668be21a966f7f2f50,
title = "Evaluation of RF localization for deep brain implants",
abstract = "This paper presents first results of an advanced study based on numerical simulations concerning the applicability of RF techniques for the localization of functional medical probes located in the ventricular system of the brain. The functional implant contains a dipole antenna that transmits short pulses in the frequency range of 3.1 < f[GHz] < 5.1. The localization concept is based on angle of arrival determination at several reference stations located around the skull of the phantom. The investigation of three test cases shows that the accuracy is significantly impaired if the spherical wave front is overlapped by a guided wave in the bone layer of the skull. Reasonable localization accuracy can be achieved if the reference stations are located at positions where the skull bone is thin.",
author = "D. Manteuffel and Hoeher, {P. A.} and M. Mehdorn",
year = "2011",
language = "English",
isbn = "9788882020743",
series = "Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011",
pages = "2372--2376",
booktitle = "Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011",
note = "5th European Conference on Antennas and Propagation, EUCAP 2011 ; Conference date: 10-04-2011 Through 15-04-2011",

}

Download

TY - GEN

T1 - Evaluation of RF localization for deep brain implants

AU - Manteuffel, D.

AU - Hoeher, P. A.

AU - Mehdorn, M.

PY - 2011

Y1 - 2011

N2 - This paper presents first results of an advanced study based on numerical simulations concerning the applicability of RF techniques for the localization of functional medical probes located in the ventricular system of the brain. The functional implant contains a dipole antenna that transmits short pulses in the frequency range of 3.1 < f[GHz] < 5.1. The localization concept is based on angle of arrival determination at several reference stations located around the skull of the phantom. The investigation of three test cases shows that the accuracy is significantly impaired if the spherical wave front is overlapped by a guided wave in the bone layer of the skull. Reasonable localization accuracy can be achieved if the reference stations are located at positions where the skull bone is thin.

AB - This paper presents first results of an advanced study based on numerical simulations concerning the applicability of RF techniques for the localization of functional medical probes located in the ventricular system of the brain. The functional implant contains a dipole antenna that transmits short pulses in the frequency range of 3.1 < f[GHz] < 5.1. The localization concept is based on angle of arrival determination at several reference stations located around the skull of the phantom. The investigation of three test cases shows that the accuracy is significantly impaired if the spherical wave front is overlapped by a guided wave in the bone layer of the skull. Reasonable localization accuracy can be achieved if the reference stations are located at positions where the skull bone is thin.

UR - http://www.scopus.com/inward/record.url?scp=79959652598&partnerID=8YFLogxK

M3 - Conference contribution

AN - SCOPUS:79959652598

SN - 9788882020743

T3 - Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011

SP - 2372

EP - 2376

BT - Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011

T2 - 5th European Conference on Antennas and Propagation, EUCAP 2011

Y2 - 10 April 2011 through 15 April 2011

ER -