Abstract

Purpose Precision skull surgery requires specialized instrumentation to satisfy demanding requirements in cochlear array implantation, deep brain stimulation electrode placement, and related applications.Aminiaturized reconfigurable parallel kinematic mechanism which can be directly mounted on a patient's skull was designed, built, and tested for precision skull surgery. Methods A Stewart-Gough platform is attached to a patient's skull so no optical tracking affecting the overall accuracy in keyhole surgery is required. Six bone anchors comprising the mechanism base joints are implanted at positions with sufficient skull thickness. Since no fixation frame is required, intervention planning flexibility is increased. The centers of the spherical shaped bone anchors can be localized accurately in the image space. An implicit registration to the physical space is performed by assembling the kinematics. Registration error is minimized compared to common optical tracker-based approaches. Due to the reconfigurable mechanism, an optimization of the hexapod's configuration is needed to maximize accuracy and mechanical stability during the incision. Mathematical simulation was conducted to estimate system performance. Results Simulation results revealed significant improvement in accuracy and stability when exploiting the redundant degrees of freedom and the implemented reconfigurability. Inaccurate localization of base points in the image data set was identified as the main source of error. A first prototype with passive prismatic actuators, i.e. micrometer calipers, was successfully built. Conclusions A head-mounted parallel kinematic device for high precision skull surgerywas developed that provides submillimetric accuracy in straight-line incisions. The system offers enhanced flexibility due to the absence of a rigid fixation frame.

ASJC Scopus Sachgebiete

Zitieren

Design and Analysis of a Head-Mounted Parallel-Kinematic Device for Skull Surgery. / Kobler, Jan Philipp; Kotlarski, Jens; Öltjen, Julian et al.
in: International journal of computer assisted radiology and surgery, Jahrgang 7, Nr. 1, 01.2012, S. 137-149.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Kobler JP, Kotlarski J, Öltjen J, Baron S, Ortmaier T. Design and Analysis of a Head-Mounted Parallel-Kinematic Device for Skull Surgery. International journal of computer assisted radiology and surgery. 2012 Jan;7(1):137-149. doi: 10.1007/s11548-011-0619-8
Kobler, Jan Philipp ; Kotlarski, Jens ; Öltjen, Julian et al. / Design and Analysis of a Head-Mounted Parallel-Kinematic Device for Skull Surgery. in: International journal of computer assisted radiology and surgery. 2012 ; Jahrgang 7, Nr. 1. S. 137-149.
Download
@article{a3f261440fea4363a2f7acdc8c7fa73f,
title = "Design and Analysis of a Head-Mounted Parallel-Kinematic Device for Skull Surgery",
abstract = "Purpose Precision skull surgery requires specialized instrumentation to satisfy demanding requirements in cochlear array implantation, deep brain stimulation electrode placement, and related applications.Aminiaturized reconfigurable parallel kinematic mechanism which can be directly mounted on a patient's skull was designed, built, and tested for precision skull surgery. Methods A Stewart-Gough platform is attached to a patient's skull so no optical tracking affecting the overall accuracy in keyhole surgery is required. Six bone anchors comprising the mechanism base joints are implanted at positions with sufficient skull thickness. Since no fixation frame is required, intervention planning flexibility is increased. The centers of the spherical shaped bone anchors can be localized accurately in the image space. An implicit registration to the physical space is performed by assembling the kinematics. Registration error is minimized compared to common optical tracker-based approaches. Due to the reconfigurable mechanism, an optimization of the hexapod's configuration is needed to maximize accuracy and mechanical stability during the incision. Mathematical simulation was conducted to estimate system performance. Results Simulation results revealed significant improvement in accuracy and stability when exploiting the redundant degrees of freedom and the implemented reconfigurability. Inaccurate localization of base points in the image data set was identified as the main source of error. A first prototype with passive prismatic actuators, i.e. micrometer calipers, was successfully built. Conclusions A head-mounted parallel kinematic device for high precision skull surgerywas developed that provides submillimetric accuracy in straight-line incisions. The system offers enhanced flexibility due to the absence of a rigid fixation frame.",
keywords = "Cochlear implant surgery, High precision skull surgery, Minimally invasive surgery, Robot-assisted surgery",
author = "Kobler, {Jan Philipp} and Jens Kotlarski and Julian {\"O}ltjen and Stephan Baron and Tobias Ortmaier",
year = "2012",
month = jan,
doi = "10.1007/s11548-011-0619-8",
language = "English",
volume = "7",
pages = "137--149",
journal = "International journal of computer assisted radiology and surgery",
issn = "1861-6410",
publisher = "Springer Verlag",
number = "1",

}

Download

TY - JOUR

T1 - Design and Analysis of a Head-Mounted Parallel-Kinematic Device for Skull Surgery

AU - Kobler, Jan Philipp

AU - Kotlarski, Jens

AU - Öltjen, Julian

AU - Baron, Stephan

AU - Ortmaier, Tobias

PY - 2012/1

Y1 - 2012/1

N2 - Purpose Precision skull surgery requires specialized instrumentation to satisfy demanding requirements in cochlear array implantation, deep brain stimulation electrode placement, and related applications.Aminiaturized reconfigurable parallel kinematic mechanism which can be directly mounted on a patient's skull was designed, built, and tested for precision skull surgery. Methods A Stewart-Gough platform is attached to a patient's skull so no optical tracking affecting the overall accuracy in keyhole surgery is required. Six bone anchors comprising the mechanism base joints are implanted at positions with sufficient skull thickness. Since no fixation frame is required, intervention planning flexibility is increased. The centers of the spherical shaped bone anchors can be localized accurately in the image space. An implicit registration to the physical space is performed by assembling the kinematics. Registration error is minimized compared to common optical tracker-based approaches. Due to the reconfigurable mechanism, an optimization of the hexapod's configuration is needed to maximize accuracy and mechanical stability during the incision. Mathematical simulation was conducted to estimate system performance. Results Simulation results revealed significant improvement in accuracy and stability when exploiting the redundant degrees of freedom and the implemented reconfigurability. Inaccurate localization of base points in the image data set was identified as the main source of error. A first prototype with passive prismatic actuators, i.e. micrometer calipers, was successfully built. Conclusions A head-mounted parallel kinematic device for high precision skull surgerywas developed that provides submillimetric accuracy in straight-line incisions. The system offers enhanced flexibility due to the absence of a rigid fixation frame.

AB - Purpose Precision skull surgery requires specialized instrumentation to satisfy demanding requirements in cochlear array implantation, deep brain stimulation electrode placement, and related applications.Aminiaturized reconfigurable parallel kinematic mechanism which can be directly mounted on a patient's skull was designed, built, and tested for precision skull surgery. Methods A Stewart-Gough platform is attached to a patient's skull so no optical tracking affecting the overall accuracy in keyhole surgery is required. Six bone anchors comprising the mechanism base joints are implanted at positions with sufficient skull thickness. Since no fixation frame is required, intervention planning flexibility is increased. The centers of the spherical shaped bone anchors can be localized accurately in the image space. An implicit registration to the physical space is performed by assembling the kinematics. Registration error is minimized compared to common optical tracker-based approaches. Due to the reconfigurable mechanism, an optimization of the hexapod's configuration is needed to maximize accuracy and mechanical stability during the incision. Mathematical simulation was conducted to estimate system performance. Results Simulation results revealed significant improvement in accuracy and stability when exploiting the redundant degrees of freedom and the implemented reconfigurability. Inaccurate localization of base points in the image data set was identified as the main source of error. A first prototype with passive prismatic actuators, i.e. micrometer calipers, was successfully built. Conclusions A head-mounted parallel kinematic device for high precision skull surgerywas developed that provides submillimetric accuracy in straight-line incisions. The system offers enhanced flexibility due to the absence of a rigid fixation frame.

KW - Cochlear implant surgery

KW - High precision skull surgery

KW - Minimally invasive surgery

KW - Robot-assisted surgery

UR - http://www.scopus.com/inward/record.url?scp=84857636877&partnerID=8YFLogxK

U2 - 10.1007/s11548-011-0619-8

DO - 10.1007/s11548-011-0619-8

M3 - Article

AN - SCOPUS:84857636877

VL - 7

SP - 137

EP - 149

JO - International journal of computer assisted radiology and surgery

JF - International journal of computer assisted radiology and surgery

SN - 1861-6410

IS - 1

ER -