Details
Originalsprache | Englisch |
---|---|
Aufsatznummer | 130 |
Fachzeitschrift | BMC veterinary research |
Jahrgang | 13 |
Ausgabenummer | 1 |
Publikationsstatus | Veröffentlicht - 11 Mai 2017 |
Abstract
Background: Aseptic loosening due to bone remodelling processes after total hip replacement is one common cause for revision surgery. In human medicine, dual-energy X-ray absorptiometry (DEXA) is the gold standard for quantitative evaluation of bone mineral density, whereas in veterinary medicine conventional radiography is used for follow-up studies. Recently, a method has been described using digital X-ray images for quantitative assessment of grey scale values of bone contrast. Therefore, the aim of the present study was to evaluate the correlation of bone mineral density (BMD) measured by DEXA with grey scale values (GV) measured in digital X-ray images (RX50, RX66) ex vivo. Results: The measured GV in the chosen X-ray settings showed on average a good correlation (r = 0.61) to the measured BMD with DEXA. Correlation between the two X-ray settings was very good (r = 0.81). For comparisons among regions of interests (ROIs) a difference of 8.2% was found to be statistically significant, whereas in the case of RX50 and RX66 differences of 5.3% and 4.1% were found to be statistically significant. Conclusions: Results indicate that measuring absolute changes in bone mineral density might be possible using digital radiography. Not all significant differences between ROIs detectable with DEXA can be displayed in the X-ray images because of the lower sensitivity of the radiographs. However, direct comparison of grey scale values of the periprosthetic femur in one individual patient during the follow-up period, in order to predict bone remodelling processes, should be possible, but with a lesser sensitivity than with DEXA. It is important that the same X-ray settings are chosen for each patient for follow-up studies.
ASJC Scopus Sachgebiete
- Veterinärmedizin (insg.)
- Allgemeine Veterinärmedizin
Ziele für nachhaltige Entwicklung
Zitieren
- Standard
- Harvard
- Apa
- Vancouver
- BibTex
- RIS
in: BMC veterinary research, Jahrgang 13, Nr. 1, 130, 11.05.2017.
Publikation: Beitrag in Fachzeitschrift › Artikel › Forschung › Peer-Review
}
TY - JOUR
T1 - Comparative measurements of bone mineral density and bone contrast values in canine femora using dual-energy X-ray absorptiometry and conventional digital radiography
AU - Lucas, K.
AU - Nolte, I.
AU - Galindo-Zamora, V.
AU - Lerch, M.
AU - Stukenborg-Colsman, C.
AU - Behrens, B. A.
AU - Bouguecha, A.
AU - Betancur, S.
AU - Almohallami, A.
AU - Wefstaedt, P.
N1 - Funding information: This study was performed in association with the subproject D6 of the collaborative research center 599: “Sustainable degradable and permanent implants out of metallic and ceramic materials”. The authors wish to thank the German Research Foundation (DFG) for its financial support.
PY - 2017/5/11
Y1 - 2017/5/11
N2 - Background: Aseptic loosening due to bone remodelling processes after total hip replacement is one common cause for revision surgery. In human medicine, dual-energy X-ray absorptiometry (DEXA) is the gold standard for quantitative evaluation of bone mineral density, whereas in veterinary medicine conventional radiography is used for follow-up studies. Recently, a method has been described using digital X-ray images for quantitative assessment of grey scale values of bone contrast. Therefore, the aim of the present study was to evaluate the correlation of bone mineral density (BMD) measured by DEXA with grey scale values (GV) measured in digital X-ray images (RX50, RX66) ex vivo. Results: The measured GV in the chosen X-ray settings showed on average a good correlation (r = 0.61) to the measured BMD with DEXA. Correlation between the two X-ray settings was very good (r = 0.81). For comparisons among regions of interests (ROIs) a difference of 8.2% was found to be statistically significant, whereas in the case of RX50 and RX66 differences of 5.3% and 4.1% were found to be statistically significant. Conclusions: Results indicate that measuring absolute changes in bone mineral density might be possible using digital radiography. Not all significant differences between ROIs detectable with DEXA can be displayed in the X-ray images because of the lower sensitivity of the radiographs. However, direct comparison of grey scale values of the periprosthetic femur in one individual patient during the follow-up period, in order to predict bone remodelling processes, should be possible, but with a lesser sensitivity than with DEXA. It is important that the same X-ray settings are chosen for each patient for follow-up studies.
AB - Background: Aseptic loosening due to bone remodelling processes after total hip replacement is one common cause for revision surgery. In human medicine, dual-energy X-ray absorptiometry (DEXA) is the gold standard for quantitative evaluation of bone mineral density, whereas in veterinary medicine conventional radiography is used for follow-up studies. Recently, a method has been described using digital X-ray images for quantitative assessment of grey scale values of bone contrast. Therefore, the aim of the present study was to evaluate the correlation of bone mineral density (BMD) measured by DEXA with grey scale values (GV) measured in digital X-ray images (RX50, RX66) ex vivo. Results: The measured GV in the chosen X-ray settings showed on average a good correlation (r = 0.61) to the measured BMD with DEXA. Correlation between the two X-ray settings was very good (r = 0.81). For comparisons among regions of interests (ROIs) a difference of 8.2% was found to be statistically significant, whereas in the case of RX50 and RX66 differences of 5.3% and 4.1% were found to be statistically significant. Conclusions: Results indicate that measuring absolute changes in bone mineral density might be possible using digital radiography. Not all significant differences between ROIs detectable with DEXA can be displayed in the X-ray images because of the lower sensitivity of the radiographs. However, direct comparison of grey scale values of the periprosthetic femur in one individual patient during the follow-up period, in order to predict bone remodelling processes, should be possible, but with a lesser sensitivity than with DEXA. It is important that the same X-ray settings are chosen for each patient for follow-up studies.
KW - Bone mineral density
KW - Bone remodelling
KW - Canine
KW - DEXA
KW - Digital radiography
KW - Femur
UR - http://www.scopus.com/inward/record.url?scp=85019127579&partnerID=8YFLogxK
U2 - 10.1186/s12917-017-1047-y
DO - 10.1186/s12917-017-1047-y
M3 - Article
C2 - 28490330
AN - SCOPUS:85019127579
VL - 13
JO - BMC veterinary research
JF - BMC veterinary research
SN - 1746-6148
IS - 1
M1 - 130
ER -