Details
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 185-192 |
Seitenumfang | 8 |
Fachzeitschrift | Acta Biotechnologica |
Jahrgang | 16 |
Ausgabenummer | 2-3 |
Publikationsstatus | Veröffentlicht - 1996 |
Abstract
In this article, attention is focussed on an on-line sampling device for bioreactors and its characterization regarding sterility and response time. The integration of this device into analytical systems for biotechnology (FIA with biosensors) resulted in on-line analysis systems for bioprocess control. The new ESIP version of an in situ sampling system for bioprocess analysis produced by the EPPENDORF-NETHELER-HINZ GmbH, Germany, had a short response time of 8 min (99%), which was determined by means of conductivity measurement after an increase of the medium conductivity due to the gradual addition of KCl. Effectiveness and reliability of the module were tested by bubble point measurement resulting in a bubble point pressure of 2.1 bars. The sampling probe was tested successfully for use in a broad variety of microorganisms and cultivations.
ASJC Scopus Sachgebiete
- Biochemie, Genetik und Molekularbiologie (insg.)
- Biotechnologie
- Chemische Verfahrenstechnik (insg.)
- Bioengineering
- Immunologie und Mikrobiologie (insg.)
- Angewandte Mikrobiologie und Biotechnologie
Zitieren
- Standard
- Harvard
- Apa
- Vancouver
- BibTex
- RIS
in: Acta Biotechnologica, Jahrgang 16, Nr. 2-3, 1996, S. 185-192.
Publikation: Beitrag in Fachzeitschrift › Artikel › Forschung › Peer-Review
}
TY - JOUR
T1 - A New Version of an In situ Sampling System for Bioprocess Analysis
AU - Hilmer, J. M.
AU - Scheper, Th
PY - 1996
Y1 - 1996
N2 - In this article, attention is focussed on an on-line sampling device for bioreactors and its characterization regarding sterility and response time. The integration of this device into analytical systems for biotechnology (FIA with biosensors) resulted in on-line analysis systems for bioprocess control. The new ESIP version of an in situ sampling system for bioprocess analysis produced by the EPPENDORF-NETHELER-HINZ GmbH, Germany, had a short response time of 8 min (99%), which was determined by means of conductivity measurement after an increase of the medium conductivity due to the gradual addition of KCl. Effectiveness and reliability of the module were tested by bubble point measurement resulting in a bubble point pressure of 2.1 bars. The sampling probe was tested successfully for use in a broad variety of microorganisms and cultivations.
AB - In this article, attention is focussed on an on-line sampling device for bioreactors and its characterization regarding sterility and response time. The integration of this device into analytical systems for biotechnology (FIA with biosensors) resulted in on-line analysis systems for bioprocess control. The new ESIP version of an in situ sampling system for bioprocess analysis produced by the EPPENDORF-NETHELER-HINZ GmbH, Germany, had a short response time of 8 min (99%), which was determined by means of conductivity measurement after an increase of the medium conductivity due to the gradual addition of KCl. Effectiveness and reliability of the module were tested by bubble point measurement resulting in a bubble point pressure of 2.1 bars. The sampling probe was tested successfully for use in a broad variety of microorganisms and cultivations.
UR - http://www.scopus.com/inward/record.url?scp=0029852271&partnerID=8YFLogxK
U2 - 10.1002/abio.370160216
DO - 10.1002/abio.370160216
M3 - Article
AN - SCOPUS:0029852271
VL - 16
SP - 185
EP - 192
JO - Acta Biotechnologica
JF - Acta Biotechnologica
SN - 0138-4988
IS - 2-3
ER -